Lecture 4 - Circulatory System
Lecture 4 - Circulatory System
Systole-
The period of contraction in cardiac cycle
Pressure and Volume Changes during the
Cardiac Cycle
• Atrial Systole- contraction of atria
• During atrial systole, which lasts about 0.1 sec, the atria are
contracting. At the same time, the ventricles are relax
• Depolarization of the SA node causes atrial depolarization.
• Atrial depolarization causes atrial systole
• As the atria contract, they exert pressure on the blood within, which
forces blood through the open AV valves into the ventricles.
• Ventricular Systole – contraction of the ventricles
• During ventricular systole, which lasts about 0.3 sec, the ventricles are
contracting.
• At the same time, the atria are relaxed in atrial diastole.
• About 0.05 seconds, both the SL (semilunar) and AV valves are closed.
This is the period of Isovolumetric contraction.
• During this interval, cardiac muscle fibers are not yet shortening. Thus,
the muscle contraction is isometric.
• Because all four valves are closed, ventricular volume remains the same
(isovolumic)
• Ventricular systole cont…
• Continued contraction of the ventricles causes pressure inside the
chambers to rise sharply.
• At this point, ejection of blood from the heart begins.
• The left ventricle ejects about 70 mL of blood into the aorta and the
right ventricle ejects the same volume of blood into the pulmonary
trunk.
• Relaxation Period
• During the relaxation period, which lasts about 0.4 sec, the atria and
the ventricles are both relaxed.
• As the heart beats faster and faster, the relaxation period becomes
shorter and shorter, whereas the durations of atrial systole and
ventricular systole shorten only slightly
Relaxation period cont…
• As the ventricles relax, pressure within the chambers falls, and blood
in the aorta and pulmonary trunk begins to flow backward toward the
the ventricles.
• Backflowing blood catches in the valve cusps and closes the SL
valves.
• After the SL valves close, there is a brief interval when ventricular
blood volume does not change because all four valves are closed.
• This is the period of isovolumetric relaxation.
Heart Sounds
• Reason- Oscillation of blood back and forth between the walls of the
ventricles initiated by inrushing blood from the atria.
Fourth Heart Sound
• It is composed of
- P wave
-QRS complex
- T wave
Reasons for different waves
P wave-
Caused by electrical potentials generated when the atria depolarize before atrial contractions
begin
This cause a slight rise in the atrial pressure curve immediately after P curve
After 0.16 seconds , the QRS waves appear
QRS complex
Caused by when the ventricles depolarize before contraction
During this process -
Depolarization wave spreads through the ventricles
Causes the ventricular pressure
The QRS complex begins slightly before the onset of ventricular systole
Each peripheral tissue of the body controls its own local blood flow
All local tissue flows combine
This flow will lead to the right atrium
The heart, automatically, pumps this incoming blood into the arteries and
lets them flow around the circuit again
Frank Starling Mechanism of the heart
• The intrinsic ability of the heart to adapt to increasing volumes of inflowing
blood
• " Within physiologic limits, the heart pumps all the blood that returns to it by
the way of the veins"
• Greater the heart muscle is stretched during filling; the force of contraction will
be high and the quantity of blood pump into the aorta will be high
Autonomic Nerve System
• Control of the heart by the sympathetic and parasympathetic nerves
• The amount of blood pump each minute can be increased more than
100 per cent via these sympathetic nerves
• The blood pump output can be decreased to low as 0 by
parasympathetic stimulation
Mechanisms of excitation of the heart by
sympathetic nerves
• Sympathetic nerve stimulation
Increase the force of heart contraction to double by increasing the volume of
blood pumped and increasing the ejection pressure
This can increase the maximum cardiac output as much as 2-fold to 3-fold
• After that, the heart starts to beat 20 to 40 beats per minute as long
SV= EDV-ESV
• End-Diastolic Volume (EDV) is the volume of blood in the right
and/or left ventricle at end load or filling in (diastole) or the amount
of blood in the ventricles just before systole.
• Greater EDVs cause greater distention of the ventricle
• End-Systolic Volume (ESV)
The volume of blood in the left or right ventricle at the end of the
systolic ejection phase immediately before the beginning of diastole or
ventricular filling
Factors affecting the Stroke Volume
• Preload
• Contractility
• Afterload
Preload
• It is the degree of myocardial distension prior to shortening
• An intrinsic property of myocardial cells is that the force of their
contraction depends on the length to which they are stretched:
• The greater the stretch (within certain limits), the greater the force of
contraction.
• An increase in the distension of the ventricle will therefore result in an
increase in the force of contraction, which will increase cardiac output.
• Preload can be increased by increasing the EDV by getting more venous
return
• Increasing ventricular filling time can increase the heart rate
• Frank–Starling law of the heart !!!
Contractility ( Increase in the force of contraction)
•Increased contractility of the heart muscle, resulting in increased cardiac
output.
• Location of measurement
The standard location for blood pressure measurement is the brachial
artery
Brachial Artery
Auscultatory and Oscillometric methods
• Auscultatory method – Involves listening to arterial sounds
• Oscillometric method - This detects variations in pressure oscillations due to
arterial wall movement.
• Manual BP measurement devices require the user to inflate the upper-arm cuff to
clog the brachial artery
• Then listen to the Korotkoff sounds through a stethoscope while the cuff is slowly
deflated.
• When the cuff is slowly deflated, five different sound phases can be heard:
• Phase I – a thud;
• Phase II – a blowing or swishing noise;
• Auscultatory gap – in some patients, the sounds disappear for a short period;
• Phase III – a softer thud than in phase I;
• Phase IV – a disappearing blowing noise;
• Phase V – silence: all sounds disappear
• Practically, the systolic reading is when the Korotkoff sounds are first heard, and
• Devices that are generally used for manual BP measurement include:
• Aneroid sphygmomanometer –
This replaces the mercury manometer
Requires use of a stethoscope
• Electronic sphygmomanometer –
This battery-powered device replaces the mercury manometer with a pressure
sensor and electronic display.
The display may be numerical, or a circular or linear bar graph. No stethoscope
is needed.
Automated electronic BP devices
• The arm should be positioned at heart level: if it is lower than the heart, this
can lead to overestimation, while being above the level of the heart can lead
to an underestimation.
• If seated, the patient should not cross their legs as this can lead to an
increase in BP.
Cuff size
• Miscuffing – particularly using a cuff that is too small – can lead to inaccurate
readings
If a cuff is too small, the BP will be overestimated and, if it is too big, the BP
will be underestimated.