Principles of CT
Principles of CT
Definition
CAT SCAN = Computed Axial Tomography
Gantry
Operator’s
Console
Table
CT Generations
1st generation Scanner
Characteristics…….
- single pencil beam
- single detector
- together translate through
180 steps and
then rotate 1o at a time
through 360o
- Scan time 3 – 5 min/slice
2 nd
Generation Scanner
• Narrow fan beam
• Small curved array of
detector
• Together they translate
and then rotate,
• But requires fewer
angular steps to produce
image
• Reduces scan time to 20
sec/slice
3rd Generation Scanner
• Most commonly used scanner
• Wide fan beam
• Larger curvilinear array of
hundreds of detectors
• Which do not translate but
simply rotate continuously
through 360o around the
patient scan time reduce to
1sec/slice
• Tube and detectors always in a
geometrical relationship
allowing for better pre detector
collimation
• which makes the image subject
to ring artifacts.
4 Generation scanner
th
512
512
Pixel Voxel
ACQUIRING THE DATA
The X ray from the tube is attenuated by the
contents within the Voxel
This attenuated ray then falls on the detector
The detector measures the logarithm of the
intensity of the radiation falling upon it.
Beam Hardening
The use of High KV (typically 120KV)
reduces:
-Patients dose
-Beam hardening by the patients
-the efficiency of the detector
-The contrast of the image
Beam hardening….
- It however increases scatter
-necessitating collimation in front of the
detectors.
This beam hardening effect results in the CT
Number of each kind of tissue decreasing along
the path of the ray
but the reconstruction process assumes the CT
numbers to be constant.
This beam hardening effect is
accommodated by the computer using a Beam hardening
algorithm.
Recently improved beam hardening algorithms have
allowed
- lighter filtration
- greater tube output
- and a shorter scan time.
- Also to compensate for diminishing patient thickness
toward the edges of the fan beam a bow tie filter is used
which is thicker along the two parallel edges than the
center.
The bowtie filter
minimizes
surface dose up to 50%
and reduces X-ray
scattering effects.
1974 CT image resolution
Characteristics
- the couch moves continuously at a steady
speed
- The tube and detectors make a number of
revolutions around the patient
- The tube receives power supply through
slip rings
- The detectors send their signals by radio
Spiral scanning
Spiral scanning
Pitch
This is defined as the distance in ( mm) moved by
the table per 3600 rotation divided by the slice
thickness also in (mm)
• Increasing the table speed, is like stretching the
spring.
• Therefore if the pitch increases
– it speeds up scanning
- Reduces dose
- But resolution may be lost
Above a pitch of 2:1 there are gaps in the volume
being scanned and artifacts may arise
Pitch Definition
Patient Table dist per 360° rotation, in relation to the slice thickness
Gantry Rotation
Table distance
Example:
Table dist = 20 mm
Slice thick. = 10 mm
Pitch =20/10 =2
Note the following:
Rotation of tube per sec (rotation/sec)
e.g 1 rotation/sec
Couch moves ; 10mm/sec
Slice thickness; 10mm
Pitch is 10mm / 10mm 1:1
- Data is acquired in the form of a continuous
ribbon of contiguous slices
- Each ribbon slanted like the turns of a spiral ring
- Data is reconstructed as a series of vertical slices
5 Generation Scanner
th
Allows user to extract a region or
anatomical feature
from a data set (referred to as volume
segmentation)
and display the result as one or more 3D
objects.
08/29/23 JDM Rev 0 44
3D Parameters
Thresholding
- Extracts a range of voxels which represent a
specific tissue or object
Scalpel
- Allows user to define a volume or region of
interest
Rendering modes
- Define how the 3D object will be displayed these
include
i. Surface shading
Treats a 3D object as if it were fully
08/29/23 opaque JDM Rev 0 45
3D Imaging
RaySum
- The pixel value is the sum of the voxel values along the
projection. Result is similar to conventional radiography.
51 44 32 45 58 22 252
51 44 32 45 58 22 42