Chapter 1 Introduction in Biomedical Sensors
Chapter 1 Introduction in Biomedical Sensors
Note –These lectures taken from text book and different internet resources and
modified by Dr. Noman AL Naggar
Plan
Introduction & concepts_
Biomedical Sensor technology_
Sensor types_
Measurements systems_
Sensor Error Sources_
Sensor Terminology and What’s requirements to Sensors_
3-Slide
Biomedical sensors
Introduction - concept
A sensor (also called detector) is a converter that measures a •
physical quantity and converts it into a signal which can be read by an
.observer or by an today mostly electronic instrument
Sensor / Transducer -Integrated with other parts to “read” out the signal (electrically, •
optically, chemically)
Some are used in vivo to perform continuous, invasive or non-invasive •
monitoring of critical physiological variables
pressure, flow, concentration of gas –
Some are used in vitro to help clinicians in various diagnostic •
procedures
electrolytes, enzymes, metabolites in blood –
concept
in vivo: inside a living body (human or animal) •
ex vivo: outside the living body •
in vitro: in a test tube •
in situ: right in the place where reactions happen (could be in the cells, tissue, test • 4
Transducers
Transducer
a device that converts a primary form of energy into a-
corresponding signal with a different energy form
Primary Energy Forms: mechanical, thermal, electromagnetic, optical,
.chemical, etc
take form of a sensor or an actuator-
Sensor (e.g., thermometer)
a device that detects/measures a signal or stimulus-
”acquires information from the “real world-
Actuator (e.g., heater)
a device that generates a signal or stimulus-
sensor intelligent
real
feedback
world
actuator system
sensor
internet
photodiode
light I
10-Slide
Biomedical Sensor technology
The Role of Sensors in BME
Sensors
Biomedical Sensor technology
Trends in biomedical sensors
12-Slide
Signal Acquisition
Medical Instrumentation typically entails
monitoring a signal off the body which is
analog, converting it to an electrical
signal, and digitizing it to be analyzed by
.the computer
Signal Acquisition
Physiological output Signals
Three types of output signal
:Self-generating (active) transducers-1
The electrical signal output of –
transducer is generated from another
.form of input energy
:Tandem transducers-3
The original input signal energy is –
converted to a final output of electrical
energy through two or three effects or
.conversions in tandem 14-Slide
Signal Acquisition
15
:Types of Sensors
Electrodes: acquire an electrical signal
20-Slide
Measurements systems
Optical Measurements are used in biomedical sensors
Optical Measurement
.Optical systems are widely used in medical diagnosis, especially in clinical-chemistry lab –
Application example: Blood or tissue sample analysis, oxygen saturation of hemoglobin, –
.cardiac output
Radiation sources: Tungsten lamp, Arc discharges, LEDs, LASERS –
: Figure
a) General block diagram of an optical
.instrument
b) Highest efficiency is obtained by
using an intense lamp, lenses to gather
and focus the light on the sample in the
.cuvette, and a sensitive detector
c) Solidstate lamps and detectors may
.simplify the system
21-Slide
Measurements systems
Optical Measurements are used in biomedical sensors
22
Measurements systems
Optical Measurements are used in biomedical sensors
Absorption/Fluorescence
Different dyes show peaks of different values at different
concentrations when the absorbance or excitation is plotted against
.wavelength
Phenol Red is a pH sensitive reversible dye whose relative
absorbance (indicated by ratio of green and red light transmitted) is
.used to measure pH
.HPTS is an irreversible fluorescent dye used to measure pH
Similarly, there are fluorescent dyes which can be used to measure
.O2 and CO2 levels
23-Slide
Connecting Sensors to Microcontrollers
sensor µC keypad
signal timing
sensor memory display
:Analog instrument
Output Output
Input Input
Input
Sensitivity Error
The pressure transducer discussed above may have an actual sensitivity
.of 7.8 mV/V/mm Hg instead of 10 mV/V/mm Hg
:Sensor Terminology
Range = Maximum and Minimum values of applied
.parameter that can be measured
If an instrument can read up to 200
mmHg and the actual reading is 250
mmHg then you have exceeded the
.range of the instrument
:Sensor Terminology
Dynamic Range: total range of sensor for minimum to
maximum. if your instrument can measure from -10V to
+10 V your dynamic range is 20V
Precision = Degree of reproducibility denoted as the
range of one standard deviation σ
Resolution = smallest detectable incremental change of
input parameter that can be detected
What’s requirements to Sensors
-2
Accuracy Sensor Terminology
Xi Xo
What’s requirements to Sensors
Offset Error -3 Sensor Terminology
Output Output
Input Input
Offset Error
Zero offset error
The difference between the actual output value and the specified
.output value under some particular set of conditions
What’s requirements to Sensors
Linearity -4 Sensor Terminology
Input
What’s requirements to Sensors
Hysteresis -5 Sensor Terminology
Hysteresis = measurement of how sensor changes
.with input parameter based on direction of change
Output = F(x)
P
F2
F1 Input = x
B
Q
What’s requirements to Sensors
Response Time -6 Sensor Terminology
Ton Time
ResponseWhat’s
-6 requirements to Sensors
Time Sensor Terminology
Ton Time
Dynamic-7
Linearity Sensor Terminology
Measure of a sensor’s ability to follow rapid changes in the input
parameters. Difference between solid and dashed curves is the non-
.linearity as depicted by the higher order x terms
Where F(X) is the output signal, while the X terms represent the input
.parameter and its harmonics, and K is an offset constant
F(x)* = ax + bx2+cxK4+ . . . +K F(x)* = ax + bx3+cxK5+ . . . +K
x + x +
Output = m Output = m
( x) ( x)
F(x) F F(x) F
K K
Input X Input X
Frequency Response of Ideal and Practical System
In Ideal: When you look at the frequency response of an
.instrument, ideally you want a wideband flat frequency response
Av Av = Vo/Vi
1.0
Av Av = Vo/Vi
1.0
0.707
FL FH
Frequency (w) radians per second
Examples of Filters
Ideal Filter has sharp cutoffs and a flat pass
band
Most filters attenuate upper and lower
frequencies
Other filters attenuate upper and lower
frequencies and are not flat in the pass
band
Sensor Calibration
Sensors can exhibit non-ideal effects
offset: nominal output ≠ nominal parameter value
nonlinearity: output not linear with parameter changes
cross parameter sensitivity: secondary output variation with, e.g.,
temperature
Calibration = adjusting output to match parameter
analog signal conditioning
7.000
digital calibration
6.000
,T = a + bV +cV2
5.000
1001
1010
;T= temperature; V=sensor voltage
Frequency (MHz)
4.000
1001
Compensation
1111
2.000
48-Slide
48.Sensors p