Electro Analytical Methods Class Pp..
Electro Analytical Methods Class Pp..
Electroanalytical Methods
1 Sisay K.
Outline
2
INTRODUCTION TO ELECTROCHEMISTRY
POTENTIOMETRY
POLAROGRAPHY
CONDUCTOMETRY
COULOMETRY
Sisay K.
Objectives
3
Sisay K.
Learning outcomes
4
electrochemical methods
Sisay K.
INTRODUCTION
5
Definition:
Sisay K.
INTRODUCTION…
6
Classification:
The simplest division is between
Sisay K.
INTRODUCTION…
7 Sisay K.
INTRODUCTION…
8
Terminology:
Electrolyte: A substance present in solution which is at least
partly in the form of charged species (ions).
Cations:
Anions:
Sisay K.
INTRODUCTION…
9
Red Cat
An ox
Sisay K.
INTRODUCTION…
11
Terminology…
Half-cell:
the anode or the cathode compartment and
all reactions that occur at that particular electrode.
Half-reaction: the redox reaction that occurs in one half-cell
Interface (Junction):
the location where two distinct phases come in contact with each other: Salt
Bridge
Reduction and oxidation
Oxidizing Agent and reducing agent Sisay K.
INTRODUCTION…
12
Electro neutrality:
Suppose, that we immerse a piece of zinc metal in pure water.
Sisay K.
INTRODUCTION…
13
Sisay K.
INTRODUCTION…
14
Sisay K.
INTRODUCTION…
15
Sisay K.
INTRODUCTION…
16
Sisay K.
INTRODUCTION…
17
e.g. batteries
b. Electrolytic cells
• Electrolysis is used to drive an oxidation-reduction and
therefore is non-spontaneous.
Electrical enrgy to chemical energy….eg electroplating
Sisay K.
INTRODUCTION…
18
cathode
Sisay K.
INTRODUCTION…
Differences
19
Sisay K.
20
POTENTIOMETRY
Sisay K.
Potentiometry
21
Sisay K.
Potentiometry…
22
Sisay K.
Potentiometry…
23
Uses symbols to identify different phases and that lists the composition of each phase.
Vertical slash(|)--- to identify a boundary between two phases
Shorthand cell notations begin with the anode and continue to the cathode
Sisay K.
25
Sisay K.
26
Example
Write the reactions occurring at the anode and the cathode for
the poten-tiometric electrochemical cell with the following
shorthand notation.
Sisay K.
Potentiometry…
27
Q is given by
2.
3.
4.
5.
Sisay K.
Potentiometry…
30
Sisay K.
Potentiometry…
31
Exercise
The Cu2+ ion concentration in a copper-silver electrochemical
cell is 0.1M. If Eo(Ag+/Ag) = 0.8V, Eo(Cu2+/Cu) = 0.34V, and
Cell potential (at 25oC) = 0.422V, find the silver ion
concentration.
Sisay K.
Potentiometry…
32
Types of Electrodes
1. Reference electrodes:
o Electrode with known constant potential
o Potentiometric electrochemical cells are constructed such that
one of the half-cells provides a known reference potential
The potential of the other half-cell indicates the analyte’s
concentration.
o By convention
Sisay K.
Potentiometry…
33
Properties of an ideal RE
It must provide a stable potential so that any change in E
cell
is attributed to the indicator electrode, and, therefore, to a
change in the analyte’s concentration.
ceases.
Sisay K.
Potentiometry…
34
Sisay K.
Potentiometry…
35
Sisay K.
Potentiometry…
36
Sisay K.
Potentiometry…
37
Sisay K.
Potentiometry…
38
Sisay K.
Has a potential of +0.2444 V at 25 °C .
39 Sisay K.
40
Sisay K.
Potentiometry…
41
Types of Electrodes…
2. Indicator electrodes
The reference electrode represents half of the complete system
for Potentiometric measurements.
ions).
Can be either Metallic or membrane electrodes
Sisay K.
Potentiometry…
43
Sisay K.
Potentiometry…
44
Sisay K.
Potentiometry…
45
Sisay K.
Potentiometry…
46
Sisay K.
Potentiometry…
47
Sisay K.
Potentiometry…
48
When the potential of an electrode of the first kind responds to
the potential of another ion
that is in equilibrium with Mn+, it is called an electrode of
the second kind.
Sisay K.
Potentiometry…
49
Sisay K.
Potentiometry…
50
Sisay K.
Potentiometry…
51
Sisay K.
Potentiometry…
52
Sisay K.
53 Sisay K.
Potentiometry…
54
o Its use is dependent upon the fact that when a glass membrane is
immersed in a solution,
o a potential is developed which is a linear function of
hydrogen ion concentration of the solution.
Sisay K.
Potentiometry…
56
E= E* - 0.0592 pH
Sisay K.
Potentiometry…
57
Sisay K.
Potentiometry…
58
Sisay K.
Potentiometry…
59
Sisay K.
Potentiometry…
60
Sisay K.
Potentiometry…
61
Sisay K.
Potentiometry…
62
Applications of Potentiometry
pH determination
Potentiometric titrations
Sisay K.
63
Polarography
Sisay K.
Polarography…
64
Include
Polarography
Coulometry and
Voltammetry
Sisay K.
Polarography…
65
Sisay K.
Polarography…
66
Thus,
a three electrode system and
a supporting electrolyte
Sisay K.
Polarography…
Instrumentation – Three electrodes
in solution containing analyte
Working electrode
Reference electrode
Auxiliary or counter
electrode
Working electrode:
microelectrode whose potential is
varied with time
68 Sisay K.
Polarography…
69
Sisay K.
Polarography…
70
Sisay K.
Polarography…
71
Real world sample solutions and standards which have very low
analyte concentrations often do not have sufficient electrolyte
content- supporting electrolyte
Sisay K.
Polarography…
72
Polarography
Is a technique which utilizes a special non-stationary electrode
Sisay K.
73 Sisay K.
Polarography…
74
The drop forming at the capillary tip grows, and it tears away
when at a time after the start of formation it has attained a certain
mass
After its detachment a new drop starts to form and grow, and the
cycle is repeated
Sisay K.
Polarography…
75
Advantages
Sisay K.
Polarography…
76
II. The decay of the current to small values with time due to the
diffusion – limited transportation of analyte to the electrode
surface when the solution is not stirred
• The solution is stirred in the case of the DME each time the
drop fall
Sisay K.
Polarography…
77
Disadvantages
It is usually used as a cathode
Diffusioncurrent (id),
Half-wave potential (E1/2),
Back ground (Residual) current, the limiting current
and
Decomposition potential
Sisay K.
79 Sisay K.
Polarography…
80
Sisay K.
Polarography…
81
substance.
Ilkovic examined the various factors which govern the diffusion
current
It is
a characteristic of the nature of the reactive material
independent on concentration
depend on pH of the medium, type of supporting
electrolyte and type of electrode
Sisay K.
Polarography…
83
Sisay K.
Polarography…
84
Polarographic maxima
Inhibit the accurate measurement of the diffusion current.
Are distortions of the current manifested by large peak of
current immediately beyond the half-wave potential.
Sisay K.
Polarography…
85
Example:
۩ Curve A is that for copper ions in 0.1M potassium hydrogen
citrate solution
۩ Curve B is the same polarogram in the presence of 0.005
Sisay K.
Polarography…
86
Oxygen waves
Dissolved oxygen is readily reduced at various electrodes; an
aqueous solution saturated with air exhibits two distinct waves
attributable to this element.
The first results from the reduction of oxygen to peroxide.
O2(g)+2H++2e= H2O2
Sisay K.
Polarography…
87
Oxygen waves…
The second corresponds to the further reduction of the
hydrogen peroxide.
H2O2+2H++2e = 2H2O
Sisay K.
Polarography…
88
Oxygen waves…
As a result, during an experiment, a suitable solution of the
electroactive substance under examination is freed-from
oxygen by:-
bubbling oxygen free nitrogen through for several minutes
and
blanketing the solution with nitrogen.
Sisay K.
Polarography…
89
Applications
COULOMETRY
Sisay K.
Coulometry…
91
Sisay K.
Coulometry …
93
Sisay K.
Coulometry …
94
Sisay K.
Coulometry …
95
A. Controlled-Potential Coulometry
The potential is selected so that the desired oxidation or
reduction reaction goes to completion
The current is proportional to the analyte’s concentration.
As electrolysis progresses the analyte’s concentration
decreases, as does the current.
Sisay K.
Coulometry …
96
A. Controlled-Potential…
The resulting current-versus-time profile for controlled-
potential coulometry, which also is known as potentiostatic
coulometry, is shown in Figure below
Sisay K.
Figure: Current versus time for a controlled-potential
coulometric analysis
The measured current is shown by the red curve. The
integrated area under the curve, shown in blue, is the total
charge(Q)
97 Sisay K.
Coulometry …
98
A. Controlled-Potential…
From the figure current decreases continuously throughout
electrolysis.
Sisay K.
Coulometry …
99
A. Controlled-Potential…
o For this reason controlled-potential coulometry is carried out
o in small-volume electrochemical cells
o using electrodes with large surface areas and
o with high stirring rates
o A quantitative electrolysis typically requires approximately
30–60 min, although shorter or longer times are possible.
Sisay K.
Coulometry …
102
Sisay K.
Coulometry …
103
A. Controlled-Potential…
The other essential feature of instrumentation for controlled-
potential coulometry is a means of determining the total
charge passed during electrolysis.
Sisay K.
Coulometry …
104
The total charge at the end of the electrolysis then can be read
directly from a digital readout or from a plot of Current versus
time
Sisay K.
Coulometry …
105
B. Controlled-current coulometry
A second approach to coulometry is to use a constant current
in place of a constant potential
Controlled-current coulometry, also known as amperostatic
coulometry or coulometric titrimetry
Has two advantages over controlled-potential coulometry
Sisay K.
Coulometry …
106
Sisay K.
Coulometry …
107
B. Controlled-current coulometry…
There is a need for a method of determining when the analyte
has been exhaustively electrolyzed
In controlled-potential coulometry this is signaled by a decrease
in the current to a constant background or residual current
Sisay K.
Coulometry …
108
B. Controlled-current coulometry…
The same end points that are used in redox titrimetry such as
visual indicators, and potentiometric and conductometric
measurements, may be used to signal the end point of a
controlled-current coulometric analysis
Sisay K.
Coulometry …
109
Instrumentation:
Sisay K.
Coulometry …
110
B. Controlled-current coulometry…
Digital clocks provide a more accurate measurement of time.
Sisay K.
111 Sisay K.
Applications of coulometry
112
Coulometric titrtions
Quantitative calculations
Example 1
To determine the purity of a sample of Na2S2O3, a sample is titrated
coulometrically using I- as a mediator and I3– as the titrant. A
sample weighing 0.1342 g is transferred to a 100-mL volumetric
flask and diluted to volume with distilled water. A 10.00-mL portion
is transferred to an electrochemical cell along with 25 mL of 1 M
KI, 75 mL of a pH 7.0 phosphate buffer, and several drops of a
starch indicator solution. Electrolysis at a constant current of 36.45
mA requires 221.8 s to reach the starch indicator end-point.
Determine the sample’s purity. (Ans 98.73% w/w)
Sisay K.
Applications of coulometry…
113
Practice Exercise 2
To analyze a brass alloy, a 0.442-g sample is dissolved in acid and
Sisay K.
CONDUCTOMETRY
114 Sisay K.
Conductometry
115
General Considerations
An electrolytic solution contains free ions.
One of the most important characteristics of electrolyte
solution is their capability to carry electric current
The electrolyte conductance is possible through movement of
positive and negative ions, which originate through dissociation
of the electrolyte.
Sisay K.
Conductometry…
116
Electrolytic conduction
exhibited principally by molten salts and by aqueous
solutions of electrolytes.
Sisay K.
Conductometry…
117
Conductometry
done by applying an electric field between two electrodes
The principle of electrolytic conduction is best illustrated by
reference to an electrolytic cell such as that shown in below for
the electrolysis of molten NaCl b/n inert electrodes
Sisay K.
Conductometry…
118
Sisay K.
Conductometry…
120
interionic attraction,
solvation of ions,
temperature and
viscosity of solvents
Sisay K.
Conductometry…
121
Sisay K.
Conductometry…
122
Sisay K.
Conductometry…
123
Sisay K.
Conductometry…
124
Sisay K.
Conductometry…
125
e.g.
NaCl behaves as strong electrolyte whereas acetic acid as a weak
electrolyte in water.
On the other hand, when dissolved in ammonia both NaCl and
acetic acid show comparable behavior towards electricity.
Sisay K.
Conductometry…
126
Conductance of solutions
Sisay K.
Conductometry…
127
Sisay K.
Conductometry…
128
Conductance of solutions…
Sisay K.
Conductometry…
129
Conductance of solutions…
Since a solution is a three-dimensional conductor, the exact
resistance will depend on the spacing (l) and area (A) of the
electrodes.
The resistance of a solution in such situation is directly
proportional to the distance between the electrodes and
inversely proportional to the electrode surface area.
Sisay K.
(rho)-Resistivity(Specific resistance)
Conductometry…
130
Conductance of solutions…
The reciprocal of resistivity is the conductivity, k(Kappa)
(Specific conductance) which is the conductance of one meter
cube of substance and has the units Ω -1 m-1
But if rho is measured in ohm cm, then k will be measured Ω -1
c m-1.
Most of the data in the literature are expressed in terms of k
measured in units Ω -1 cm-1
Sisay K.
Conductometry…
131
Conductance of solutions…
The expression for the conductance is therefore
and , therefore
Cell constant: for a given cell, l and A are constant, and the
quantity (L/A) is called the cell constant(K).
Sisay K.
Conductometry…
132
Sisay K.
Conductometry…
133
Sisay K.
Conductometry…
134
Sisay K.
Conductometry…
135
Sisay K.
Conductometry…
136
Sisay K.
Conductometry…
137
Molar Conductivity
The conductance of an electrolytic solution depends mainly on
the type of ions present,
speed with which the ions move in the solution and
their concentration.
Sisay K.
Conductometry…
138
Molar Conductivity…
When a solution of elctrolyte is diluted, the conductance will
decrease.
Sisay K.
Conductometry…
139
Molar Conductivity…
This is due to
a decrease in inter-ionic effects for strong electrolytes
and
an increase in the degree of dissociation for weak
electrolytes.
Sisay K.
Conductometry…
140
Molar Conductivity…
The molar conductivity (Л) (Lambda) of an electrolyte is defined
Л = 1000k/C = k. 1000 V
Sisay K.
Conductometry…
141
Molar Conductivity…
At infinite dilution, the ions are quite far apart, the inter-ionic
attractions are almost absent and each ion moves completely
independent of co-ions.
The molar conductivity then approaches a limiting value at
infinite dilution and represents the conducting power of one
mole of the electrolyte when it is completely slit up into ions.
It is denoted by Л
8
.
Sisay K.
Conductometry…
142
Molar Conductivity…
Conclusion
in case of weak electrolytes, the increase in molar conductivity
with dilution is mainly due to
Measurement of conductance
The conductance of a solution can be determined by
measuring the resistance
For measuring resistance, the Wheatstone bridge principle is
used.
It works on the principle of obtaining balance between two
arms with the help of a balance indicator (e.g. a
galvanometer) at the condition of potential being equal.
Sisay K.
Conductometry…
144
A DC Wheatstone bridge
circuit
Sisay K.
Conductometry…
145
Sisay K.
Conductometry…
146
Sisay K.
Conductometry…
147
EB = ED or i1R1 = i2R2
Sisay K.
Conductometry…
148
Sisay K.
Conductometry…
149
Applications
- conductimetric titrations (Reading A.)
Sisay K.
150
The End
Sisay K.
QUIZ
151
1. What is conductometry?
Conductometry means measuring the conductivity of an
electrolyte or molten salts
Sisay K.