Module 4 Biomedical
Module 4 Biomedical
Therapeutic Equipments
AKHIL KUMAR S
ASSISTANT PROFESSOR
ECE
MLMCE
AKHIL KUMAR S, AP, ECE
CARDIAC PACEMAKERS
AKHIL KUMAR S, AP, ECE
Internal Pacemakers:
In this entire system is inside the body.
Permanently implanted in the body whose SA node failed to
function properly.
The system is implanted with the pulse generator placed in a
surgically formed pocket below the right or left cavicle.
Internal leads connected to electrodes that directly contact
surface of myocardium
Pulse generator must be self contained with a power source
capable of continuously operating the unit for a period of
years.
AKHIL KUMAR S, AP, ECE
External Pacemakers
AKHIL KUMAR S, AP, ECE
POWER SUPPLY:
It is required to supply energy to the pacemaker
Primary or secondary batteries are used as power source
For long life lithium batteries are used
Sometimes external power sources can be used for
implantable pacemakers.
OSCILLATOR:
AKHIL KUMAR S, AP, ECE
Pacemaker electrodes
Synchronous Pacemaker
AKHIL KUMAR S, AP, ECE
In cases of complete heart block where the atria are able to
depolarize but the impulse fails to depolarize the ventricles,
atrial synchronous pacing may be used.
Here the pulse generator is connected through wires and
electrodes to both the atria and the Ventricles.
The atrial electrode couples atrial impulses to the pulse
generator, which then emits impulses to stimulate the
ventricles via the ventricular electrode.
In this way, the heart is paced at the same rate as the
natural pacemaker.
When the SA node rate changes because of sympathetic
neuronal control, the ventricle will change its rate accordingly
but not above some maximum rate (about 125 per minute).
CARDIAC DEFIBRILLATORS
AKHIL KUMAR S, AP, ECE
1. AC Defibrillator
2. Capacitive Discharge DC Defibrillator
components:
The defibrillator itself (AID), the lead system, and the
programmer recorder/monitor (PRM).
The AID houses the power source, sensing, defibrillation,
pacing, and telemetric communication system.
The leads system provides physical and electrical connection
between the defibrillator and the heart tissue.
The PRM communicates with the implanted AID and allows
the physician to view status information and modify the
function of the device as needed
Programmer Recorder/Monitor (PRM)
AKHIL KUMAR S, AP, ECE
Controlled mode
In this mode the breathing is controlled by an automatically
timing system which is usually provided for patients who
cannot breath on their own.
Pressure control (PC)
Volume control (VC)
Assisted mode/Supported mode
In this mode patients own spontaneous attempt to breath in,
causes ventilator to cycle on during inspiration
Continuous positive airway pressure (CPAP)
Pressure support
Assist Control mode/ Combined mode
AKHIL KUMAR S, AP, ECE
Inspiration
It is an active movement
The diaphragm moves downwards
The outside air goes into lungs
Expiration
It is a passive movement
Return of the diaphragm to normal
Air (more CO2 goes out from lung
Tidal volume/lung volume
The volume of air/O2 for one breath.
Unit is ml
Breath rate
Number of breaths for one minute.
Minute volume
AKHIL KUMAR S, AP, ECE
mechanical ventilator.
It consists of a microprocessor with RAM, EPROM, A/D
converter and a CRT controller.
The input signals to the microprocessor are obtained from a
CO2 analyser, a lung machine, gas analyser, oxygen
consumption monitor and a servo ventilator.
The proper controlling signals are delivered to the servo
ventilator so as to get correct ventilation adjustment in
response to a patient‘s metabolism.
HEART-LUNG MACHINE
AKHIL KUMAR S, AP, ECE
Pumps:
The pumps are designed to minimize the damage to blood
cells and effective in pumping within physiological range.
Membrane Oxygenator
It imitates the function of lungs.
Membrane oxygenator are more common now a days.
Here O2 & CO2 Exchange takes place.
Gas exchange take place by the process of diffusion across a
thin membrane separating blood and gas made of highly
permeable silicon rubber or microporous polypropylene,
Teflon & polyacrylamide.
Heat Exchanger
AKHIL KUMAR S, AP, ECE
is sent.
The developed heat produces a wedge shaped narrow
cutting of the tissue on the surface.
By increasing the current level, deeper level cutting of
the tissue takes place.
Normally continuous RF current is used for cutting
COAGULATION
When the electrode is kept near the skin, high frequency
current is sent through the tissue in the form of bursts
and heating it locally so that it coagulates from inside.
The concurrent use of continuous RF current for cutting and
a RF wave burst for coagulation is called Haemostasis.
FULGURATION
This is called Fulguration‘ in which the electrode is held near
AKHIL KUMAR S, AP, ECE