Basic CT training : X-ray II

TSG

CT

Generation of X-rays

15 sheets
x-ray2 *C

 Copyright Toshiba 1997 1


Basic CT training : X-ray II
TSG

CT

Functions of X-ray generator

MAINS IN TO XRAY TUBE

Low Voltage Increase High Voltage: kV

High Current Decrease Low Current: mA

AC Rectify DC

kV mA time

Operator Controls
 Copyright Toshiba 1997 2
Basic CT training : X-ray II
TSG

CT

Basic circuit of X-ray generator


AUTO HIGH VOLTAGE X-RAY
TRANSFORMER TIMER TRANSFORMER RECTIFIER TUBE

High Voltage
Low Voltage
Low Current
High Current
DC
AC

CATHODE

kV s mA
HEATER
POWER
LINE

 Copyright Toshiba 1997 3


Basic CT training : X-ray II
TSG

CT

H.V. Transformer + Rectifier


360 M

+ kV
+ mA 48 K

50

- mA

- kV

 Copyright Toshiba 1997 4


Basic CT training : X-ray II
TSG

CT

Generation of X-rays
vacuum

ANODE KATHODE
100 mA

100 mA 100 mA
+ -
30 - 150 kV

 Copyright Toshiba 1997 5


Basic CT training : X-ray II
TSG

CT

Metal Section X-ray Tube


Metal Primary
Center Section Electrons

ANODE 90 mA CATHODE 100 mA

Secondary
Electrons
Center Section Lead

10 mA
90 mA 100 mA

 Copyright Toshiba 1997 6


Basic CT training : X-ray II
TSG

CT

Metal Section X-ray Tube (2 focus)

Filament for Small Focus

ANODE CATHODE Common

Filament for Large Focus

ANODE ANODE ANODE

Filament for Small Focus

Common
Focal
Large Small Spot
Focus Focus Filament for Large Focus

 Copyright Toshiba 1997 7


Basic CT training : X-ray II
TSG

CT

Metal Small Large


Focus Focus
Section Focal spot

X-ray Tube
(2 focus) Same Slit

 Copyright Toshiba 1997 8


Basic CT training : X-ray II
TSG

CT

Metal Section X-ray Tube (2 focus)


Small Focus Large Focus

– Lower power input + High power input


(e.g.: 120kV 200mA) (e.g.: 120kV 300mA)

+ For small objects – Not for thin slices.


(thin slices)

 Copyright Toshiba 1997 9


Basic CT training : X-ray II
TSG

CT

Heat producedmAin X-ray tube (1/3)


kV Time

Heat
Produced

Cooling Unit

OIL

Housing

Anode Focal Spot

 Copyright Toshiba 1997 10


Basic CT training : X-ray II
TSG

CT

Heat produced in X-ray tube (2/3)


Focal Spot
mA
kV Time
radiation
conduction

Heat Anode
Produced
radiation

Housing
Cooling Unit

AIR

convection

 Copyright Toshiba 1997 11


Basic CT training : X-ray II
TSG

CT

Heat produced in X-ray tube (3/3)


HEAT HEAT
Temperature
4000o

3000o

2000o

1000o

0o

Small heat capacity


Large heat capacity
 Copyright Toshiba 1997 12
Basic CT training : X-ray II
TSG

CT

OverLoad Protection (OLP)


• OLP will protect the tube. It will prevent
exposure if the tube is too hot.

• OLP is calculated by the input kV, mA and


time. This is expressed in Heat Units (HU).
kHU = kV mA  sec  1.4

• The X-ray tube specification will state the


maximum amount of HU the tube can
handle. Also a cooling curve is specified.
 Copyright Toshiba 1997 13
Basic CT training : X-ray II
TSG

CT

Anode heating and cooling curves

 Copyright Toshiba 1997 14


Basic CT training : X-ray II
TSG

CT

Tube protection
Cooling Unit

Temperature
OIL
Switch

Oil Flow
Switch

In case the temperature switch or/and the flow switch


detect an abnormality, X-ray generation is not possible.

 Copyright Toshiba 1997 15