Multi Intro2
Multi Intro2
A-serie MultiSlice
CT
Introduction
Toshiba 2002 1
Multislice Helical CT
Instantaneous Volume
Toshiba 2002
Imaging
2
Development of CT Scanner
Single Breath Hold Scanner
High-speed 3D scanner
Multislice CT
Half-second 0.5s CT
Sub-second 0.75s CT
Real-time CT 1994
Helical CT 1990
Conventional CT
Toshiba 2002 3
A CLEAR VISION FOR THE
FUTURE
256
elements
256 Slice
64 Slice
Advanced Multi-II
16 slice technology
Advanced Multi
1234
High Res.
1234 Mode
1 2 3 4
Z-direction 1 2 3 4
Detector
1 2 3 4
elements
High Speed
Switches 1 2 3 4 Mode
DAS DAS DAS DAS
DAS for for for for Slice Thickness
1st slice 2nd slice 3rd slice 4th slice
for MultiSlice 0.5mm to8.0mm @ 8 slices/sec
Toshiba 2002 6
Multislice Cone-beam
Tomography (MUSCOT)
Reconstruction Algorithm
• Reconstruction algorithm for Helical Scan
• Patent applied for, paper published in Medical Physics,
April 1998
• Markedly enhanced image quality with cone beam
geometry
• Significant reduction of partial volume artifact
• New reconstruction algorithms allow choice of arbitrary
slice profile (Reconstruction slice thickness).
Toshiba 2002 7
Images of ball
phantoms
Single - slice MultiSlice
Helical scan Helical scan
Toshiba 2002 8
Multislice Helical CT
4-slice data acquisition per rotation
30cm 30cm
10mm-slice 10mm-slice
30sec 4sec
Toshiba 2002 10
Multislice Helical CT
Clinical Benefit - Improved Z-Resolution
Single-slice Helical Scan Multislice Helical Scan
5 3 8
Only those three parameters are special in
Multi-slice CT.
Toshiba 2002 12
Isotropic Image
• High 3D image quality Same resolution in 3 directions
• With the Dual detector, with 0.5 mm slice and 896 channels it
is possible to get an isotropic volume image
• Reduction of the “partial volume” effect
0.35 mm
0.5 mm
0 .3 0.3
5m m 5m
m m m m
m 35
.35 0.
0
a b c
Toshiba 2002 14
Specifications
Aquilion Asteion Asteion/Dual
0.32 (229 ), 0.5, 0.75, 0.5 (229 ), 0.75, 1.0, 1.5, 2.0 and 3.0
Scan Speed
1.0, 1.5, 2.0 and 3.0 sec. sec.
0.5 mm x4 0.5 mm x2
1.0 mm x4 1.0 mm x2
2.0 mm x4 2.0 mm x2
3.0 mm x4 3.0 mm x2
Slice thickness’
4.0 mm x4 4.0 mm x2
5.0 mm x4 5.0 mm x2
8.0 mm x4 8.0 mm x2
10 mm x2 10 mm x2
Reconstruction time 0.5 s (0.5s scan) 0.5 (0.75s scan) 1.0 s (0.75s scan)
Raw Data Storage 4000 rotations
Image Data Storage 16000 images 8000 images
High Voltage 60kW 36*, 48, 54* or 60kW*
Generator (80, 100, 120 , 135kV) (80, 100, 120 , 135kV)
HeliCool (4.0 Mhu)
X-ray tube MegaCool (7.5 Mhu)
6.5Mhu*
* = optional
Toshiba 2002 15
Features
Fast MPR
3D
Standard Sure Start
Real-time Helical Scan
Multiview
CT-Fluoroscopy
Cardiac gated scanning
Cardiac gated reconstruction
DICOM MWM
DICOM PPS
DICOM Storage SCP
Options DICOM Q&R SCU / SCP
Bone Mineral Study
Dental application
Fly-through
Pediatric scanning
Extend Raw Data Disk
Extend Image Data Disk
Toshiba 2002 16
Anet (Plus) option
Toshiba 2002 18
Option: Cardiac
Scanning:
•Scan & ScanProspective Gating
•Half-scan acquisition
0.32 sec
Table slide
Toshiba 2002 19
Cardiac Scanning
Coronary artery
Toshiba 2002 20
Cardiac Scanning:
Retrospective Gating
•Helical Scan
•Half-scan acquisition
Axial
Sagital
Coronal
Toshiba 2002 23
Service related
• Tube alignment: same as Single slice.
– F/B: also possible without film, using SSMD.
• Calibration: same as Single slice.
– L and LL phantoms: plastic.
– 1 water scan.
– 1 air scan on
all rotation speeds.
Toshiba 2002 24