BMEG 3320: Biomedical Imaging Magnetic Resonance Imaging
BMEG 3320: Biomedical Imaging Magnetic Resonance Imaging
2
https://www.youtube.com/watch?v=3nIXRPuFK5U
MRI scanner
4
Typical floor layout of MRI system
5
MRI Hardware System
• Magnet room:
• Main Magnet: Generate B0 field
• Gradient coils: Generate gradient
of magnetic field in the direction
of B0
• RF Coils: Generate Radiofrequency
waveforms to excite the protons
and detect the signal from the
protons after excitation
• Equipment room:
• System electronics
• Control room:
• Viewing Console
6
Main Magnet
7
Main and shielding coils of a The six primary coils (1–6) and two shield coils (7,
superconducting magnet before being 8) mounted on an aluminium alloy bobbin
enclosed in a cryostat.
8
The invisible force
https://youtu.be/6BBx8BwLhqg 9
Gradient Coils
Helmholz coil configuration The birdcage coil is the most commonly used RF-
transmit device used in clinical MRI today http://mri-q.com/ 11
RF coils
• Receiver only RF coils
• RF transmission and reception functions have been largely separated and delegated to
two different types of coils.
• In the typical arrangement, RF-transmission is performed by a body RF-coil located just
beneath the inner wall of the scanner bore. RF-reception is delegated to one or more
surface coils placed close to the patient's body as illustrated in the diagram above.
• RF receiver chain:
Coil Decoupling → Preamplification → ADC → Demodulation → Image Processing
http://mri-q.com/
13
Put all together The essential components of
a typical clinical MRI system
"An MR scanner is a coil within a coil within a coil within a coil . . . "
Schematic cross-section
through a typical
superconducting clinical MR
scanner.
https://hackaday.io/project/182802-home-made-mri
https://tabletop.martinos.org/index.php? 15
Outline
• Hardware of MRI system
• MR Image Formation
• Gradient Echo vs. Spin Echo
• Contrast, Field of View, Resolution, and SNR in MRI
16
Data acquisition
Image reconstruction
Post-processing
Signal Acquisition
• Receive sum of signals from each spin.
18
Reconstruction
• Received signal is a sum of each waveform
• Each waveform has its own frequency
• Fourier Transform can be used to obtain each frequency component
Fourier
Transform
Position
20
Gx a linear gradient Gy a linear gradient
along x along y
Bz=B0 +xGx Bz=B0 +yGy
z
z
y
y
x
x
Bz Bz
21
• Why do we use gradient?
• Why not use other non-linear magnetic field variations?
22
MR Signal
The received signal is proportional to the vector sum of the
magnetization within the imaging volume:
𝑆 ( 𝑡 )=∫ 𝑚 ( 𝑟
𝑗 𝜑 (𝑟
⃗ ,𝑡 )
⃗ )𝑒 𝑑⃗
𝑟
𝑉
: spin density;
23
MR Signal
Remember the phase is time integral of precession frequency.
Thus =
Therefore 𝑡
𝑗 2 𝜋 𝛾 ∫ ( 𝑥 𝐺𝑥 (𝜏 )+ 𝑦 𝐺 𝑦 (𝜏 )) 𝑑 𝜏
𝑆 ( 𝑡 )=𝑒 𝑗 2 𝜋𝛾 𝐵 0𝑡
∫ 𝑚 (𝑟⃗ ) 𝑒 0
𝑑⃗
𝑟
𝑉 24
k-space
Define
𝑆 ( 𝑡 )= 𝑠 ( 𝑘𝑥 , 𝑘 𝑦 ) =∫ 𝑚 ( ⃗
𝑗 2 𝜋 (𝑥 𝑘 + 𝑦 𝑘 𝑦)
𝑟)𝑒 𝑥
⃗
𝑑𝑟
𝑉
Thus the acquired signal and the image are Fourier transform of
each other:
s kx , k y F m x, y
( k x ,k y )
25
t t
k x (t ) Gx d , k y (t ) G y d
2 0 2 0
ky
Gy
a) t=0 kx
Gx
Data Acquisition
a) t=0 b) t=T c) t=2T d) t=3T
26
2D MR imaging
ky
kx
28
Selective Excitation
29
Selective Excitation
After we apply gradient Gz along z direction, the Larmor frequency at location z is
𝜔 0 +𝛾 𝐺𝑧 𝑧
The RF pulse usually has a bandwidth , which means its rotation frequency has a range
Thus, the RF pulse will excite a slice with thickness
Position
gG
Frequency
(a)
(b)
RF Amplitude
Magnitude
Frequency Time
(c) (d)
31
2D MR imaging - Phase Encoding and Frequency
Encoding
• Phase Encoding:
• Apply gradient to impose a spatially dependent
Phase encoding phase on the signal from the precessing protons
• The gradient is turned on and off before data
acquisition
• A number of different values of the phase-
encoding gradient must be used.
• Frequency Encoding:
Frequency encoding • Apply a gradient to create a spatially dependent
precessional frequency during signal acquisition
Data acquisition
• The gradient is turned on during data acquisition
32
K-space
trajectory
Spins in
Transverse
plane Frequency
encoding
Phase
encoding
33
Pulse Sequences
• Excitation and imaging are separate.
• Pulse sequence controls:
• RF excitation
• Gradient waveforms
• Acquisition
• Reconstruction information as well.
34
2DFT - Pulse Sequence
RF
Gz
Gx
Gy
Acq.
TR, TE
Excitation Imaging 35
TE and TR
• The echo time (TE) represents the time from the center of the RF-pulse
to the center of the echo.
36
Outline
• Hardware of MRI system
• MR Image Formation
• Gradient Echo vs. Spin Echo
• Contrast, Field of View, Resolution, and SNR in MRI
37
Gradient Echo
• Undoing of phase shifts from gradient fields
𝑡
Gx
Location x1
Location x2
t
S(t) t=T t=2T
t=T t=2T
38
Spin echo
• Undoing of phase shift from field inhomogeneities and chemical shift
• Ideally, the magnetic field should be homogeneous in the MR scanner, but there are imperfections in
reality. The field inhomogeneity causes protons resonant at different frequencies even there is no
gradient turned on.
39
Four Basic Types of Signals in MRI
40
Outline
• Hardware of MRI system
• MR Image Formation
• Gradient Echo vs. Spin Echo
• Contrast, Field of View, Resolution, and SNR in MRI
41
Contrast
• Contrast is the difference in appearance of different tissues in an image.
43
T2 Contrast (long TE, long TR)
Short Echo-Time Long Echo-Time
CSF
Signal
White/Gray Matter
Time
44
T1 Contrast (short TE, short TR)
Short Repetition Long Repetition
White/Gray Matter
Signal
46
Short TR long TR long TR
Short TE long TE short TE
47
Sampling in frequency domain
Field of View corresponds to replication in
the image domain
F.T
.
Frequency Image
F.T.
https://www.youtube.com/watch?v=qn5WhQ0jIY0&feature=youtu.be
50
Tp
∆ 𝐺𝑦
Phase encoding
Gy
Gx
Frequency encoding T
Samplin Image
g
Pattern
in k-
space
Samplin
g Image
Pattern
in k-
space
52
Resolution
• Image resolution increases as higher spatial frequencies are acquired.
1 mm 2 mm 4 mm
ky ky ky
kx kx kx
53
Image Noise and SNR
Low Signal-to-Noise Ratio High Signal-to-Noise Ratio
54
Example
We want to acquire a 256x256 image with a pixel resolution of 1mm. Thus FOV is
25.6cm. We know that Gx = 40mT/m.
1. What is the sampling step (sampling rate) T ? What is the duration of the
readout gradient Ts?
2. If the duration of the phase encoding gradient Tp = Ts/10, what is the value ?
Answer:
1 1
1. 𝑇= = =2.29 𝜇 𝑠
𝛾 𝐺 𝑥 𝐹𝑂𝑉 𝑥 6
10 𝐻𝑧 −3
42.58 × ×0.256 𝑚 × 40 ∗10 /𝑚
𝑇
Duration of readout gradient Ts = 256*T = 0.586ms
1 1
∆ 𝐺 𝑦= = =1.57 mT / m
2. 𝛾 𝑇𝑝 ∙ 𝐹𝑂𝑉 𝑦 6
10 𝐻𝑧 −3
42.58 × × 0.256 𝑚 ×0.0586 × 10 𝑠𝑒𝑐
𝑇
55
Acknowledgement
• Prof. Thierry Blu
• Prof. Brian Hargreaves, Stanford University
References:
• Dwight G. Nishimura, Principles of Magnetic Resonance Imaging
• Bernstein, King, Zhou, Handbook of MRI pulse sequences
• http://mri-q.com/
56