SCHWIND ORK-CAM TreatmentTypeDecisionTree and Customization V2.2 (1)
SCHWIND ORK-CAM TreatmentTypeDecisionTree and Customization V2.2 (1)
SCHWIND ORK-CAM
2
Treatment Type Decision Tree &
(Re-) Treatment Customization
Overview
Introduction
Overview
Introduction
Treatment preparations
General: The more information you have the more individual the treatment and the better
Introduction the result could be. Minimum 3 good quality measurements per eye (of each diagnostic
device) allow a valid comparison of examinations.
Treatment
Technological view:
preparations
For an adequate and coherent analysis of the eye both a Corneal Wavefront Analyzer and an
Decision Tree criteria Ocular Wavefront Analyzer (Aberrometer) is advised for wavefront comparison and analysis.
For higher precision and better efficacy of a customized treatment a compensation of
Optical Zone selection possible cyclotorsion movements is recommendable:
SCC (Static Cyclotorsion Control)
Adjustment of laser DCC (Dynamic Cyclotorsion Control)
refraction
Equipment needed:
Comparison of
ablation depths • Corneal Wavefront Analyzer (Sirius, Scout, or Peramis)
• Ocular Wavefront Analyzer (Peramis or irx3)
• SCC from Corneal and Ocular Wavefront Analyzer
• SCC and DCC within the AMARIS
Introduction
What is necessary for perfect evaluation and planning of customized treatments?
Other recommendations:
Comparison of
ablation depths • SCHWIND Treatment Type Decision Tree consideration
• Use of the SCHWIND CAM <Comparison> module
• Advanced knowledge in CUSTOMIZED treatment planning
Introduction
Aberrometer measurements (e.g. PERAMIS)
Introduction
Improved pupil recognition of aberrometer examinations:
» Perform minimum 3 measurements per eye
Treatment
preparations » Cover the fellow eye and advise patient to look towards or through the target
» Let the patient adapt again to dark room conditions before starting the next acquisition
Decision Tree criteria
Introduction
Adjustment of laser
refraction * with use of PERAMIS
Comparison of
ablation depths
Introduction
What is necessary for perfect evaluation and planning of customized treatments?
1. Select for the best topography and wavefront maps according to the quality criteria
Introduction 2. Do a comparison of OW and CW data within SCHWIND PERAMIS directly or make use of the
SCHWIND Comparison module (after export of diagnostic patient eye data)
3. Use the ‘Export to SCHWIND’ within the SCHWIND diagnostic devices for diagnostic patient file
Treatment
preparations creation
4. Determine the best (treatment) option (AF, OW, CW, no laser treatment) for the individual
Decision Tree criteria patient, taking patient history and anamnesis into consideration SCHWIND Decision Tree
5. Make use of subjective and objective refraction information to include both refractive, corneal,
Optical Zone selection
and wavefront influences into the treatment plan: “adjusted refraction” (in OW or CW)
10. Create an SCHWIND ORK-CAM export file for treatment with AMARIS laser
Treatment preparations
Select the best measurement - PERAMIS/SIRIUS - (Keratoscope, Aberrometer, Scheimpflug,…)
Treatment
preparations
Adjustment of laser
refraction
Comparison of
ablation depths
Introduction » Analysis diameter: The bigger the analysis diameter, the more expressive the information
and the larger the OZ for wavefront comparison and treatment. A dark room is mandatory
for aberrometer measurements. Let the patient adapt to the dark room conditions for
Treatment
preparations several minutes. Min. export diameter is 5.00 mm but the larger the better.
» SCC detection: For more precision and better efficacy of a customized treatment a
Decision Tree criteria
compensation of possible cyclotorsion movements is recommendable. SCC shall be
suitable (green mark) in case the cylinder refraction >1.25D
Optical Zone selection
» Aberrometer refraction: Check power of Sph.Eq. [D] and Cylinder [D] at 4.0 mm analysis
diameter and compare to manifest (subjective) values Prefer the measurement with
Adjustment of laser most positive SEq value. Max. standard deviation (SD) should be 0.25D. In case of
refraction
astigmatism >1.25 D the max. deviation in axis should be 10°.
Comparison of
ablation depths The recommendable best ocular wavefront exam might be a compromise between the
maximum available analysis diameter, the most positive SEq, and the exam with the
lowest higher order aberrations (RMS value).
Treatment preparations
Select the best aberrometer measurement (PERAMIS Ocular Wavefront)
Introduction
Treatment
preparations
PERAMIS - Aberrometer
Decision Tree criteria
Adjustment of laser
refraction
Comparison of
ablation depths
The best measurement (quality and reproducibility criteria are met) is set to favorite.
Decide whether full pupil (øMin) is large enough for analysis and a possible OW treatment
Treatment preparations
Select the best aberrometer measurement (PERAMIS Ocular Wavefront)
Treatment
preparations
1. vs. 2. vs. 3.
Adjustment of laser
refraction
Comparison of
ablation depths
Look for the most positive refraction (to minimize accommodative effects)
Treatment preparations
Select the best aberrometer measurement (PERAMIS Ocular Wavefront)
Treatment
preparations
Adjustment of laser
refraction
Comparison of
ablation depths
Reconsider the pre-defined favorite measurement in case of high differences in HOA RMS:
ideally includes the most positive refraction AND the lowest RMS HOA at full pupil.
Treatment preparations
Select the best aberrometer measurement (PERAMIS Ocular Wavefront)
Optical Zone selection provides the refractive effect in photopic-like light conditions.
Consider the Cyl [D] + Ax value (WF Ast @ 4mm) for refractive
comparison/calculation purposes.
Adjustment of laser
refraction • The astigmatism with reference to 6.0 mm analysis diameter
Introduction » Keratoscopy quality [%]: OK if green check mark (for sufficient centration and coverage)
exists discard measurement for analysis if attention sign is shown
» Placido coverage [%]: OK if no attention signs are visible. Otherwise, perform new
Treatment measurements with eyes wide open ideally, take the highest value in %
preparations
» Section (Scheimpflug) coverage [%]: OK if no attention sign is visible. Otherwise, perform
new measurements with eyes wide open ideally, take the highest value in %
Decision Tree criteria
» SCC Check (Static Cyclotorsion Control): For more precision and better efficacy a
compensation of possible cyclotorsion movements is recommendable SCC shall be
Optical Zone selection suitable (green mark) in case the cylinder refraction >1.25D and for customized treatments
» Check Meridians 3mm (Avg) [D] max. SD (standard deviation) 0.25D
» Check Meridians 3mm (Cyl) [D] in case cylinder >1.25D the max. deviation in axis 10°
Adjustment of laser
refraction » Check CCT [μm] (Central Corneal Thickness) max. SD 10 μm
» Pupil center [mm] to check offset information: The crosses shall be inside the yellow
Comparison of
ablation depths
dotted ellipse and as close together as possible prefer the smallest offset out of
reproducible measurements
» Keratoscopy center [mm]: The crosses shall be inside the yellow dotted ellipse and as close
together as possible prefer the closest out of reproducible measurements
Treatment preparations
Select the best topographic measurement (PERAMIS/SIRIUS Corneal Wavefront)
SIRIUS - Scheimpflug
Introduction
Treatment
preparations
PERAMIS - Keratoscope
Adjustment of laser
refraction
Comparison of
ablation depths
The best measurement (quality and reproducibility criteria are met) is set to favorite.
Treatment preparations
Select the best Scheimpflug measurement (SIRIUS Corneal Wavefront)
Introduction
Treatment
preparations
Correct Scheimpflug detection is also essential for wavefront calculation processes and purposes
Treatment preparations
Select the best topographic measurement (PERAMIS/SIRIUS Corneal Wavefront)
» Homogeneity: The best topography should be homogeneous with no artefacts (e.g. eyelid,
eyelashes, tear film, stripes due to movement ). exclude unstable tearfilm and eyelid
Introduction artefacts
» Check Offset values on consistency and magnitude if the offset ≥ 0.50mm repeat
Treatment measurements and take care to cover the fellow eye and advise patient to fixate the target
preparations correctly
» Check for indication of keratoconus (KC) check KC screening and the summary of KC
Optical Zone selection indices. The individual cornea can be classified:
Adjustment of laser
refraction
Comparison of »
ablation depths
Go to Export “Refractive Export to SCHWIND” to check for placido rings, pupil & limbus:
• Check amount of placido rings Ideally ≥ 14 full Placido rings are available (minimum: 12 rings)
• Check correct limbus detection red half circles fit with limbus
Treatment preparations
Select the best topographic measurement (PERAMIS/SIRIUS Corneal Wavefront)
Introduction
Treatment
preparations
Adjustment of laser
refraction
Treatment
preparations
Adjustment of laser
refraction
Comparison of
ablation depths
Treatment preparations
Select the best topographic measurement (SIRIUS Corneal Wavefront)
Treatment
preparations
Adjustment of laser
refraction
Comparison of
ablation depths
Treatment preparations
Select the best topographic measurement (PERAMIS/SIRIUS Corneal Wavefront)
Interesting parameters “customized treatment”:
Introduction » Pupil offset:
To check repeatability of the pupil offset value and axis which is important
for corneal vertex (i.e. visual axis) treatment alignment make use of
“Statistics on acquisition” or the “Comparison” function within Phoenix SW
Treatment preparations
Select the best topographic measurement (PERAMIS/SIRIUS Corneal Wavefront)
» Topographic astigmatism:
Refer to “refraction adjustment” for an adapted (laser) refraction which is useful as starting
point in the evaluation process and the final treatment decision of complex customized cases
Treatment preparations
Select the best topography and corneal wavefront measurement (Keratron Scout)
Overview about topographies (OD/OS) and its importance - select the best image
Introduction
» SCC detection » Homogeneity » Pupil detection
Adjustment of laser
refraction
Comparison of
ablation depths
Treatment preparations
Select the best topography and corneal wavefront measurement (Keratron Scout)
Overview about topographies and its importance - select the best image
Introduction
Treatment » Homogeneity:
preparations
The best topography should be
Decision Tree criteria homogeneous with no artefacts
(e.g. stripes due to movement,
eyelid, eyelashes, tear film)
Optical Zone selection
Treatment
To guarantee the efficacy of a customized treatment a compensation
preparations of possible cyclotorsion movements is recommended
» Repeatability check:
Decision Tree criteria
• Comparison of the pictures on a mathematical way.
Optical Zone selection • Accept “good” or “very good” reports only as reliable exams
Overview about topographies and its importance - select the best image
Introduction
» Pupil offset:
• The entry of the pupil offset value aligns the aspheric correction
Adjustment of laser
of the manifest refraction to the corneal vertex whereas the
refraction compensation / correction of HOA keep linked to the pupil
center
Comparison of
ablation depths • Clinical evaluation proved that a pupil offset ≥ 0.2 mm has a
significant influence on the post-OP visual performance. Check
always the repeatability of pupil offset value and axis
To check repeatability of the pupil offset value and axis which is important for corneal vertex
(i.e. visual axis) treatment alignment compare the offset readings of the single acquisitions
Treatment preparations
Select the best topography and corneal wavefront measurement (Keratron Scout)
Overview about topographies and its importance - select the best image
Introduction
» Topographic astigmatism:
Treatment • A topographical astigmatism is calculated
preparations
based on differences in perpendicular
meridians of the keratometries
Decision Tree criteria
Overview about topographies and its importance - select the best image
Introduction
Introduction
Overview about topographies and its importance - select the best image
Adjustment of laser
refraction
Comparison of
ablation depths
The astigmatism with reference to 4.0 mm The astigmatism with reference to 6.0 mm
analysis diameter provides the refractive analysis diameter provides the refractive
effect in photopic-like light conditions effect in mesopic-to-scotopic-like light
conditions
Treatment preparations
Select the best topography and corneal wavefront measurement (Keratron Scout)
Refer to “adjusted refraction” for an adapted (laser) refraction which is useful as starting point
in the evaluation process and the final treatment decision of complex customized cases
Export of Ocular Wavefront data
Refractive export display
PERAMIS – Aberrometer
PERAMIS - TopoAberrometer
Export of Corneal Wavefront data
Refractive export display
SIRIUS - Scheimpflug
PERAMIS - Keratoscope
» Aberration-Free (AF)
Introduction
Only spherical and cylindrical refraction values are corrected. As the patient is used
to his preoperative existing aberrations he retains his normal visual perception if
Customized treatment these aberrations are kept. This procedure is suitable for patients whose vision is
preparations not impaired by pronounced higher-order aberrations.
Optical Zone selection Besides the spherical and cylindrical refraction corneal wavefront data is additionally
taken into account. This procedure is suitable for patients whose vision is impaired by
pronounced higher-order aberrations on the corneal surface without internal
Adjustment of laser
refraction
compensation.
Compensation of flap-cut
Adjustment of laser induced aberrations Yes Yes Yes
refraction (biomechanical effect)
Corneal AND
Ocular Wavefronts Always No No
< 0.25 D
OW CW IW
Comparison of OW and CW data
SCHWIND CAM Comparison Module
OW
CW
IW
After import of the two wavefront files, the information refers to the smallest common import diameter.
Take 6 mm (as default) for analysis or even the diameter which is equal to your optical zone size selection.
Decision Tree criteria
SCHWIND CAM Comparison module
» Functional Optical Zone (FOZ):
• FOZ can be selected on the referring icon on the left side of the
Introduction
Comparison module and is defined as the area where the RMS HO
value exceeds 0.375D
Customized treatment • A white ring indicates the FOZ within the wavefront map. The exact
preparations
value of FOZ is mentioned below the slider bar for optical zone (OZ) size
adjustment
Decision Tree criteria
• FOZ function might be useful to analyse the change in HOA between
pre- and post-surgery (aberration-free concept), possible night vision
Optical Zone selection problems andOrder
othersAberrations (HOA):
» Level of Higher
Adjustment of laser • The significance of HOA for selected OZ is shown via traffic light
refraction
• HO refraction with sphere, cylinder and axis is displayed and expressed in
diopters. This refraction reflects the effect of HOA at selected diameter (optical
Comparison of
ablation depths zone) - expressed in spherical and cylindrical power – that is automatically
compensated for in the SCHWIND refractive treatment The software treatment
plan with ORK-CAM tries to be on patient target by this compensation
The compensation of HOAs (sphere & cylinder) is often more powerful than the treatment of HOAs (=efficacy) themselves.
Thus, reduction in sphere/cylinder (to more positive refraction values) is typically needed in OW-/CW- treatment plans.
Decision Tree criteria
SCHWIND CAM Comparison module
Workflow following the Treatment Type Decision Tree with COMPARISON module
Introduction » If the analysis diameter (OZ) is >6.0 mm change it to 6.0 mm and press <Apply> to set
all wavefront information to a standard wavefront reference diameter
Customized treatment » Wavefront A provides data of the first imported file (e.g. ocular wavefront)
preparations
» Wavefront B provides data of the second imported file (e.g. corneal wavefront)
Decision Tree criteria » Combined maps shows the difference (A-B) of the two imported files
» By choosing the <Aberration Info> button a summary of aberrations is available
Optical Zone selection
» The clinical relevance of aberration terms
Adjustment of laser can be classified as equivalent defocus
refraction (DEq). This DEq metric is identical to the
power vector notation for low order
Comparison of
ablation depths
aberrations and allows defining optical blur
in dioptres [D]. To follow the flow chart
“Treatment Type Decision Tree” take the
RMS HO of the Aberration Info.
Depending on optical zone size selection for treatment, you may change and use the same (OZ) size for analysis (diameter)
Decision Tree criteria
SCHWIND CAM Comparison module
NO
Optical Zone selection
CW
RMS HO > 0.50D YES
Adjustment of laser NO
refraction
Complaints about quality
CDVA = 1.0
NO or night vision YES
Comparison of YES NO
ablation depths
AF Customized
Aberration-Free Treatment
Decision Tree criteria
Analysis by PERAMIS TopoAberrometry or SCHWIND CAM Comparison module
» OW RMS HO = 0.29D
Continue with next steps in case of
Introduction Customized Treatment decision: » CW RMS HO = 0.44D
Assumption:
Difference OW – CW Indication and age
Optical Zone selection RMS HO > 0.375D For IOL exchange
YES YES Patient age: 30 years
NO NO
No signs of cataract
Adjustment of laser
refraction Æ OW > Æ CW
YES
NO
Comparison of
ablation depths
CW OW NO TREATMENT
Corneal Wavefront Ocular Wavefront with Excimer Laser
Decision Tree criteria
Alternative flow chart if an aberrometer-link to SCHWIND is missing
Introduction
CW
YES RMS HO < 0.25D
NO
Customized treatment CW
preparations RMS HO > 0.50D YES
NO
Comparison of
ablation depths
AF CW NO TREATMENT
Aberration-Free Corneal Wavefront with Excimer Laser
Decision Tree criteria
Summary
Adjustment of laser
refraction Re-Treatment situation
» Whatever laser system performed the previous
Comparison of ablation take the OZ of the initial treatment into
ablation depths
account
» In case no patient history is available the OZ size of the initial d = 6.19mm
treatment might be determined with the use of <FOZ> or
estimated with topography information, e.g. measure the
diameter of the “myopic red ring” in the tangential map
Optical Zone selection
Re-treatment in Aberration-Free (AF) mode
Introduction
» Simple over- / undercorrection without visual complaints and CDVA ≥ 20/20 (feet) or 1.0
Customized treatment
(decimal)
preparations
» No aim to improve CDVA with excimer laser treatment in the individual patient eye
Decision Tree criteria » No information about “internal” eye aberrations and their possible influence on patient
vision, i.e. no aberrometer data available and test with rigid gas permeable contact lens
Optical Zone selection (RGPs) not possible
» The analysis diameter of ocular or corneal wavefront data (HOA) are inefficient or way
Adjustment of laser smaller than the scotopic pupil size, even though an Optical Zone extension by 20% max.
refraction
of the diagnostic export diameter is possible within ORK-CAM software: the peripheral
area is extended by an aspheric (AF) profile fitting as soon as wavefront data are not
Comparison of
ablation depths available anymore, i.e. the diagnostic export diameter is smaller than the selected OZ
Aberration-Free
OZ re-treatment = primary OZ
Optical Zone selection
Re-treatment in Aberration-Free (AF) mode
Hyperopia
• Reason: the periphery (= TZ & untreated area) is more
Decision Tree criteria
myopic whereas within OZ hyperopia exists
• Optimum (default OZ setting):
Optical Zone selection
6.2mm < OZ ≤ 6.3mm
Note: Red colour – Myopic defocus
» To evaluate the change in SphAb make use of the <Comparison> module within the
Customized treatment
SCHWIND CAM.
preparations
» The determination of the SphAb value refers to the OZ used (tendency a little smaller)
Decision Tree criteria and must NOT automatically related to 6.0 mm.
» In case the OZ used for primary treatment was selected larger or different than the
Optical Zone selection SCHWIND proposed OZ default, take the difference of these adjustments into account:
• Example: Primary myopia treatment with OZ = 7.0 mm (default OZ = 6.3 mm) initially
Adjustment of laser chosen because of a scotopic pupil size close to 7.0 mm
refraction
‐ Calculated OZ difference (delta between primary treatment and OZ default) is 0.7 mm
» High order aberrations (HOA) might have an influence on low order aberrations
Customized treatment (sphere and cylinder) , e.g. coma aberration may contribute to the power/axis of
preparations manifest astigmatism
» HOA reduce the precision of determination of manifest refraction because of
Decision Tree criteria
low(er) image quality (= reduced CDVA) and possibly increased depth-of-focus
(= significant amounts of spherical aberration) due to previous laser treatment
Optical Zone selection » HO refraction compensation in sphere and cylinder power (which is part of the
individual customized treatment profile) increases with bigger OZ sizes and may
Adjustment of laser significantly influence the depth/volume of the treatment profile
refraction
Comparison of In case of customized OW or CW treatments the primary Optical Zone (OZ) size of the
ablation depths AF treatment does not play a role.
Customized treatments with significant high-order aberrations (HOA) consider
typically an OZ of 7 mm and more (as long as enough corneal tissue is available)
Adjustment of laser refraction
Customized OW and CW (re-)treatments
When the Treatment Type Decision Tree (flow chart) results in a customized treatment
Introduction planning it is recommended to perform a first refraction adjustment based on the summary
of:
» Manifest refraction
Customized treatment
preparations » Aberrometer refraction @ 4mm
Adjustment of laser If no aberrometer refraction is available the autorefraction might be useful to take instead.
refraction
The topographical astigmatism is taken from Meridians @ 3mm (SIRIUS, PERAMIS) or „Best
Comparison of Fit Maloney“ (Scout).
ablation depths
The wavefront astigmatism @ 4mm and 6mm is taken from the Topographer (Corneal
Wavefront treatment) or Aberrometer (Ocular Wavefront Treatment), respectively.
Based on the SEQ of the manifest refraction and the mean of the 5 astigmatism values the
adjusted refraction (“vectorial mean”) can be calculated.
Comparison of ablation depth/volume
Difference in depth/volume: Aberration-Free (AF) vs. Customized (OW/CW) treatment plan
The SCHWIND Decision Tree supports you in making the best decision for each individual
Introduction patient. Nonetheless, patient eyes with previous corneal treatments or injuries are less
predictable than virgin corneas.
Optical Zone Size Depth* Volume Volume
Customized treatment
Myopia* Hyperopia*
preparations
[µm] [nl] [nl]
Decision Tree criteria < 5.25 mm 6-8 up to 75 up to 95
5.25 - 5.75 mm 8 - 10 90 - 125 115 - 160
Optical Zone selection 5.75 – 6.25 mm 10 - 11 125 - 185 160 - 220
6.25 – 6.75 mm 11 - 13 185 - 255 220 - 300
Adjustment of laser 6.75 – 7.25 mm 13 - 15 255 - 350 300 - 400
refraction
7.25 – 7.75 mm 15 - 17 350 - 470 400 - 530
Comparison of *corresponds to a 0.75D-AF-treatment for selected OZ (which is „deviation maximum“ of AF vs. OW/CW)
ablation depths
If the differences between manifest refraction in Aberration-Free mode vs. customized
Corneal or Ocular Wavefront are larger than these for the selected OZ, modify the
spherical and/or cylindrical component with use of <Refraction>, and make use of
<Pyramid> afterwards if even more tissue has to be saved. Otherwise, the risk of a
refractive outcome significantly different from target exists.
Comparison of ablation depth/volume
Difference in depth/volume: Aberration-Free (AF) vs. Customized (OW/CW) treatment plan
Customized treatment » Wavefront-guided treatments (OW/CW) most likely take more tissue into account
preparations
than Aberration-Free (AF) treatments with equal manifest refraction values. This
Decision Tree criteria leads to the fact that wavefront-guided treatments do have a higher potential of
over corrected outcomes
» A certain depth/volume difference is tolerable if high-order aberrations are part of
Optical Zone selection
the treatment plan but being conservative, i.e. sparing tissue for another possible
optimization, should be a major goal
Adjustment of laser » If accurate refraction (= patient target refraction) is intended, i.e. the treatment plan
refraction
has no therapeutic reason, OW/CW treatment plan should not deviate more than
±0.75D from AF-depth/volume
Comparison of
ablation depths
Customized Treatment optimization - OPTION 1: Comparison of ablation
depth/volume
CW-guided (manifest) vs. AF (manifest) vs. CW-guided (adjusted)
Example Example
Myopic Astigmatism Hyperopic Astigmatism
Re-Lift OZ (mm): 7.00 PRK OZ (mm): 7.20
CW-guided (manifest) 61 59 1911 123 89 4879
AF (manifest) 13 13 367 109 31 2840
CW-guided (adjusted) 70 68 2200 121 87 4916
Default setting
The smaller the ‚Constraints‘ selected the smaller the deviation maximum
(‚Residual‘) from ‚Current‘ refraction can or will be.
Customized Treatment: The manager function
PYRAMID: Optimization possibilities (depth/volume) in comparison
Customized Treatment: The manager function
Make use of the output from OPTION 1 or OPTION 2 (with very similar results)
The manager function gives the opportunity to individualize Ocular and Corneal Wavefront treatments.
The <Manager> function has to be used after the final adjustment of treatment parameters.
The typical order of the optimization process is from “ROUGH” (Refraction 1 st) to “FINE” (Pyramid 2nd).
Customized Treatment: The manager function
REFRACTION: Maximum Saving (depth vs. volume) with NO CONSTRAINTS selection
» Calculate the difference in depth/volume between both treatment types at equal Optical Zone (OZ)
» Look for the tolerance factor (TF) or refer to the HO Refraction- Info (out of SCHWIND CAM Comparison), each
related to the OZ selection for treatment
» Increase constraints ‘Cylinder’ only if a big impact of HO astigmatism and/or Coma aberration is present. Refer
to the HO Refraction value of the cylindrical component at selected OZ (out of SCHWIND CAM Comparison
module)
» Decrease constraints ‘Cylinder’ (close to zero) if cylinder component is repeatable (subjective, topo, wavefront)
» Change constraints ‘Axis’ only if bigger axes differences than 15° exist between subjective, topo, and adjusted
» Decrease constraints ‘Axis’ (close to zero) if cylinder power and axis is repeatable (subjective, topo, wavefront)
The tighter the constraints selected, the lesser the change (“tissue saving”) of the ‘Current’ can be
SCHWIND ORK-CAM, Main Menu in OS (V5.2)
Customized treatment – Patient with progressive keratoconus
» Example of a Main Menu that displays a CW-guided treatment (SCHWIND SIRIUS) (Trans-)PRK
proposal for a patient with progressive keratoconus, followed by CXL treatment.
Agreement of the Ablation Profile
Comparison of diagnostic device information and SCHWIND ORK-CAM ablation map
PRK
SIRIUS Tangential Anterior
and Altimetric (Asphero-
Toric) Map show very high
agreement in shape to the
ORK-CAM customized profile
: 7 mm reference diameter
72
SCHWIND ORK-CAM, Summary Page in OS (V5.2)
CW-guided (SIRIUS) TransPRK with refraction & pyramid (manager) adjustment
1. Select for the best topography and wavefront maps according to the quality criteria
2. Do a comparison of OW and CW data within SCHWIND PERAMIS directly or make use of the SCHWIND
Comparison module (after export of diagnostic patient eye data)
3. Use the ‘Export to SCHWIND’ within the SCHWIND diagnostic devices for diagnostic patient file creation
4. Determine the best (treatment) option (AF, OW, CW, no laser treatment) for the individual patient, taking
patient history and anamnesis into consideration SCHWIND Decision Tree
5. Make use of subjective and objective refraction information to include both refractive, corneal, and wavefront
influences into the treatment plan: “adjusted refraction” (in OW or CW)
6. Do a comparison of Aberration-Free (AF) depth/volume vs. Wavefront-guided (OW, CW) depth/volume at
same (pre-)selected Optical Zone (OZ) size if OW or CW is considered
7. Make use of the ORK-CAM <Manager> function after final OZ size selection: tissue saving possible via
adaptation in laser refraction “Refraction” and/or in high-order aberrations “Pyramid”
8. Do a comparison between ablation map (=laser profile) and diagnostic map(s) [topography, wavefront,
elevation, …], i.e. validate whether shape and depth/volume plan are plausible.
9. Check the correctness of SCHWIND ORK-CAM input data (“Summary Page”)
10. Create an SCHWIND ORK-CAM export file for treatment with AMARIS laser
Thank you very
much for your kind attention!
Vielen Dank für Ihre Aufmerksamkeit!
Customer Support