5 Parts With Autoloader
5 Parts With Autoloader
Mindray BC5390
• 5-part
differentiation(CBC+DIFF)
on white blood cells
• Principle: Semi-conductor
laser combined with
chemical dye method
advanced flow cytometry
• Throughput of 60
samples/ hr
Features
• 21 parameters
• 40-tubes autoloader
with random access
• Closed tube for STAT
samples
• Capability to flag
abnormal samples
• Only 33 μl sample
volume for CBC + DIFF
results
Mindray BC5390
• User friendly
software
• Storage of 100,000
patient results
Mindray BC5390
• Counting modes: Venous, Capillary, Predilute
• With available holder adapters to support 3
types of samples
Mindray BC5390 Reagents
• Diluent
• Lyse 1, 2, 3
• Maintenance reagent:
Probe cleanser
Other Technical Specs
• Communication: LAN Port
supports HL7 protocol
• Operating Environment
– Temperature: 15 °C~30°C
– Humidity: 30~85%
– Air Pressure: 70~106 kPa
• Power Requirement
– A.C.100-240V ≤ 300VA 50/60Hz
• Dimension and Weight
– Dimension (WxDxH, inches):
22.4 x 23.2 x 20.6
– Weight: ≤ 143 pounds
Dymind DH71
FEATURES
29 parameters
3D scattergram
Wide linearity
80 samples per hour
Only 20µl sampling volume for 5-part
differentiation
Dymind DH71
Fully Automatic Intelligent Technology
Visual Self-tests (Working status self-test and
reagent volume self-detection)
Automatic test re-run, with re-run rules inside
One-button error removal and automatic standby
mode function
Intelligent WBC/RBC/PLT floating threshold
technology
Excellent Data Management System
A wide range of pathological alarms and up to 20
user-defined alarm ranges
Customized templates print and help to customize
the final report according to requirements
Bi-directional LIS connection
QC and Calibration management
Dymind DH71
TECHNICAL SPECIFICATIONS
Detection Principle
• 3-Angle Semiconductor laser scatter, flow
cytometry (WBC Diff)
• Impedance method (WBC/RBC/PLT)
• Cyanide-free (HGB)
Parameters
• 29 parameters are provided including 25 report
parameters (WBC, LYM%, LYM#, NEU%, NEU#, MON
%, MON#, EOS%, EOS#, BAS%, BAS#, RBC, HGB,
HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT,
PDW, MPV, PCT, P-LCR and P-LCC) and 4 research
parameters (ALY#, ALY%, LIC#, LIC%). One 3D
scattergram, three 2D scattergram, and 3
histograms
Dymind DH71
TECHNICAL SPECIFICATIONS
Throughput
• 80 samples per hour
• Up to 90 samples per hour in CBC mode
Data Storage
• Up to 100,000 results
Reagents
• DIL-A Diluent, LYA-1 Lyse, LYA-2 Lyse, LYA-3 Lyse, CLE-P
Test Modes
• CBC and CBC-Diff
Counting Modes
• Venous blood, capillary blood and pre-diluted
Sample Volume
• 20µl
Power Requirement
• 100 – 240V, ≤250VA, 50/60Hz
Dymind DH71
TECHNICAL SPECIFICATIONS
Net Weight
• 56kg
Dimension
• 650mm (L) x 610mm (W) x 550mm (H)
Linearity Range
• WBC: 0.00 – 300.00 x 10⁹/L
• RBC: 0.00 – 8.50 x 10¹²/L
• HGB: 0 – 250 g/L
• PLT: 0 – 3000 x 10⁹/L
• HCT: 0.0 – 67.0%
Repeatability
• WBC: ≤2.0%
• RBC: ≤1.5%
• HGB: ≤1.5%
• PLT: ≤4.0%
• MCV: ≤1.0%
Reagent Information
Reagent Name Shelf Life Contents Open Life
DR diluent
LD lyse
LH lyse
ESR solution
reagent
Probe
FR dye cleanser
FD dye
18
Comparative Matrix
Specifications Mindray BC5390 Dymind DH71 Mindray BC780R Symex XN550 Beckman DxH560
Sample 33 ul 20 ul 25ul 25 ul 17 ul
volume (whole
blood)