Lecture 9 - Pacemakers and Artifical Hearts
Lecture 9 - Pacemakers and Artifical Hearts
Dr. D. Saravanakumar,
Assistant Professor, SMBS,
VIT - Chennai Campus.
Email: [email protected]
• Failure at any of these points (e.g. due to fracture in the lead or loss
of contact between the electrode and the underlying myocardium)
will result in failure of myocardial stimulation.
A) Coaxial lead body design consists of two nested coil B) In coradial design, a single coil contains the
conductors to the tip and ring electrodes, each wrapped parallel conductor strands to the tip and ring
in a separate layer of insulation. The inner coil connects electrodes, with each strand covered by a
to the tip cathode and contains a central lumen for fluoropolymer(ETFE) insulator. A surrounding layer
stylet passage; the outer coil connects to the ring of outer insulation by polyurethane covers the lead
anode.can be attached to endocardium via active/passive body.Design can allow significantly smaller (<6F)
fixing mechanisms but limited by large diameters diameters but limits the fixation mechanism.
• MRI SureScan™ technology allows the patient to be safely scanned using either
a 1.5T or 3T full body MRI.
Identify Problem
Process is iterative
– You need to repeat
various steps after Specify Criteria
testing
– Make design
changes Implement Design
based on test results
Failed designs
– Design didn’t meet Test Design
criteria
– Could be due to
Refine Design
inappropriate criteria
Implant controller,
battery and coil
Connect patient to
heart-lung machine
Cut away ventricles
Sew grafts onto atria
and arteries
Connect implants to
grafts
Remove patient from
heart-lung machine