C60 ppt
C60 ppt
C80
C60 C50
Comen Patient Monitors
(3 Generations)
STAR8000H
STAR8000B C90
2nd Generation
STAR8000A
00
New Generation
1st Generation C30
Contents
The definition of Neonate
/ 1000
(populati
on)
Problems faced in
neonatal
Most monitors can not care
be accurate for all vital signs of
neonate
Because of the special parameter range of HR, RESP, SpO2,
VITAL SIGNS\
and NIBP.
patient Neonate Adult
90-180bpm, with wide
HR range 60-100bpm
RESP 30-50/min 16-20/min
SYS: 90-140mmHg
SYS: 55-90mmHg BP DIFF: 30-50mmHg
BP DIFF: 15-20mmHg relatively high
NIBP low perfusion perfusion
close to adult, but
neonates can not
produce much heat,
which leads to low
TEMP insulation stable temperature
www. szcomen. com 94%-100%
94%-100% relatively high
COMEN - SHARE WITH THE WORLD
They only focus on the accuracy of adult parameters while pay few attention
on neonate aspect.
NICU needs
specialized
neonatal monitor!!!
Specialized measurement
technology
• ExNeo™ ECG, RESP measurement technology
• MASIMO /Nellcor SpO2 measurement technology
• Adap-Dsp™ NIBP measurement technology
• CO2 measurement technology
• Unique asphyxia self-rescue function
• Oxygen saturation monitoring function
Main parameter:
• HR range:30—250bpm
• Accuracy:±10% or 5bpm
• Arrhythmia analysis : 13 kinds of Arrhythmia analysis and
ARR self study function
• Pacemaker inhibit and detection function
• RESP rate detection range: 0~120bpm
• RESP rate detection accuracy:±1bpm
• Detection sensitivity: 0.3Ω impendence change
Masimo
Masimo SpO2
SpO2 probe
probe for
for neonates
neonates
OxiMax
OxiMax ®® SpO2
SpO2 technology
technology
LoSat™ technology
OxiMax
OxiMax ®®SPO2
SPO2 technology
technology
Solutions :
Now, the difficulties faced in the measurement of neonatal SPO2 are:low
perfusion, anti-movement. While OxiMax® technology, which is fa
mous as gold standard, can promise the accurate measurement even in low
perfusion and movement state.
Main parameter :
•Measurement range : 1% ~ 100 %
•Resloution:1%
•Low perfusion measurement range : 0.03 ~ 2%
•Oxygen saturation accuracy : ±1% ( 90 %~ 100 %); ±2% ( 70% ~ 90
%);
•PR range : 20 ~ 250bpm
•PR accuracy : ±1bpm
OxiMax ®® SpO2
OxiMax SpO2 technology
technology
Specialized
Specialized SpO2
SpO2 probe
probe for
for neona
neona
tes
tes
ap-Dsp™
ap-Dsp™ NIBP
NIBP Technology
Technology
ap-Dsp™
ap-Dsp™ NIBP
NIBP Technology
Technology
Main Parameter :
• Measurement Parameter: SBP, DBP, MAP.
• Measurement Range : SBP: 5.3 ~ 20kPa ( 40 ~ 150 mmHg), DBP :
1.3 ~ 13.3kPa ( 10 ~ 100 mmHg ), MAP: 2.7 ~ 14.7kPa ( 20 ~ 110
mmHg )
• NIBP Measurement Accuracy : ±5mmHg
• NIBP Measure Repeatability : 4mmHg
• Static Pressure Range : 0mmHg~150mmHg
• Static Pressure Measurement Accuracy : ±3mmHg
• Hardware and Software OVP, Max Pressure 150mmHg
• Unique Neonate Dedicated Mode, Prevent Misoperation, make sure the
safety of NIBP measurement of Neonate.
• Only in this way can medical personnel understand and control the
change in oxygen saturation in the incubator, thus helping the
newborns out of such risky conditions as asphyxia and “oxygen
intoxication” caused by high oxygen saturation.
Standard interface
Trend screen
OxyCRG screen
List face
Powerful software
Drug-dose calculation
Powerful software
Alarm
recall
Working with
Incubator