PART ONE Moving Parts
1
0
PART TWO
Functional Parameters
2
0
PART THREE Macro
3
FOCUS MAKES PROFESSION
Maglumi Service Software
Login password:
1
1.1 Incubator
1
1.1 Incubator
a. <Writeparam> &
<Initialize>
b. Choose Stepsize
c. Initialize
Last
Incload
Backtr
d. Incubator
Position:
Incload is
selected,
adjust Pos 1-16
FOCUS MAKES PROFESSION 6/50
0 Moving Parts
PART ONE
1
1.1 Incubator
1
1.2 Wash Trans
1
1.2 Wash Trans
1
1.3 Wash Lift
1
1.3 Wash Lift
a. <Writeparam>
& <Initialize>
b. Choose Stepsize 5
1
1.4 Chamber Trans
1
1.4 Chamber Trans
a. Disconnect cables of
PMT and the reference
LED, take away the
cover of the chamber
b. <Writeparam> &
<Initialize>
c. Load Pos
b. Adjust Pos
1
1.4 Chamber Trans
1
1.5 Chamber Lift
1
1.5 Chamber Lift
a. <Writeparam>
& <Initialize>
b. Choose
Stepsize
1
1.6 Pusher
1
1.6 Pusher
a. <Writeparam> &
<Initialize>
b. Choose Stepsize
c. Init/Backtransport: Pusher to
Backtrans
Washer: Washer to Pusher
Chamber: Pusher to Chamber
1
1.7 Pumps
1
1.7 Pumps
a. Condensate Int.
b. Soak time
c. Speed
d. Pump control
1
1.8 Components that do not need adjustment
1
1.9 Frequency and Regulation
a. Frequency: speed
of the motor
b. Regulation:
regulate the
reference point
for counting the
steps
Parameters
2
2.1 Globals
Parameters
2
2.1 Globals
Functio Basic
n parameter
buttons s
Parameters
2
2.1.1 Prime W/C
Parameters
2
2.1.2 Background
a. Background/Closed
Background/Open
Background/Transport
pos:
b. Ref.
<80 LED on:
140,000-230,000
If not, change the
value in
D/A Ref Led
Parameters
2
2.1.2 Background
Parameters
2
2.1.3 Plateau
b. Reference LED on
Parameters
2
2.1.3 Plateau
Parameters
2
2.1.4 Meas. Cuv.
b. <Start Meas.>:
“please insert cuvette
in pusher ”
c. Insert a clean cuvette
in pusher, the third
notch match the
position clip
d. Result range: 200-
1200, CV<10%.
If not, check inject
needle, starter and
clean channel.
FOCUS MAKES PROFESSION 32/50
0 Functional
PART TWO
Parameters
2
2.1.5 System Test
Parameters
2
2.1.6 ErrorCard
Rang
e
Parameters
2
2.1.6 ErrorCard
Parameters
2
2.1.6 ErrorCard
Parameters
2
2.1.6 ErrorCard
Parameters
2
2.1.7 Min Value Cuv
Parameters
2 a. Go to <Background>. Measure the “Ref. LED
2.1.7 Min Value Cuv
Parameters
2
2.1.8 RLU Factor
Parameters
2
2.1.8 RLU Factor
Parameters
2
2.2 Wash
Parameters
2
2.2 Wash
2.3 LLDParameters
2 (Liquid Level Detection)
2.3 LLDParameters
2 (Liquid Level Detection)
Choose the
position
Meaning of Delta
Z
and NoLiquid
2.3 LLDParameters
2 (Liquid Level Detection)
a. 4 tubes with
500uL distilled
Example: M2000, left pipettor
water inside
b. Put these 4 tubes
in sample area:
(T1, P1), (T1, P12),
(T12, P1), (T12,
P12)
c. Put 1 empty tube
in sample area:
(T6, P6)
d. 2 empty reagent kits
with 1ml distilled
Select
water added in Box 1
and Box 7
e. Insert the kits in
reagent area: T1 and
T15
FOCUS MAKES PROFESSION 46/50