Background: Gary Lensmeyer of the University of Wisconsin Medical Center (UWMC) is responsible
for exploring new technologies for his laboratory. During Monica Howland’s (MicroLiter) employment
with 3M she worked with Mr. Lensmeyer to evaluate and assist in development of the Empore line of
SPE membranes.
MicroLiter contracted with UMWC to evaluate Instrument Top Sample Prep (ITSP) by developing a
method that would typically be used in his lab. UWMC specializes in transplantation of organs.
Immunosuppressants, prescription pain killers, and Vitamin D were the principal high volume assays
that were run daily in his lab.
Mr. Lensmeyer chose Immunosuppressants as the method he would transfer to our ITSP.
Terms of the UWMC Contract: Mr. Lensmeyer was asked to independently evaluate ITSP to give us
good, bad and indifferent opinions relating to the device. While we thought he would have a positive
experience we did not ask him to produce data that offered only the benefit of the device.
Because of conflicts in scheduling MicroLiter’s HTC PAL, the research was conducted in two phases.
The first phase would be development of the method on the HTC PAL. The second phase would be to
run the device against current laboratory operations.
Due to the critical nature of the work being performed in his lab, the Lab Manager, Anita Iwanski,
would not allow the instrument to be attached on-line with the Agilent LC/MS that was currently
running the assay.
Mrs. Anita Iwanski, Lab Manager made it clear that there was no money available for purchase of the
system until July 2010.
The Work Product: Mr. Lensmeyer’s current method utilized Phenomenex Strata
SDB-L sorbents and we were asked to use this material for a true apples-to-apples comparison. We
asked for and obtained a sufficient quantity to begin our study. Initially 10mg bedmass weights were
prepared. Mr. Lensmeyer was given 100 devices in advance of the delivery of the instrument to try and
produce a manual method that we could program into the PAL. He was able to recreate the method,
supplied us the conditions and we wrote the automation in advance of the delivery of the instrument.
On the day we installed the instrument, Mr. Lensmeyer immediately prepared solvents and supernatants
and our first test run of the instrument prepped samples that gave Mr. Lensmeyer baseline data with
which to start. The very first sample produced data that would have been acceptable to the lab to provide
the physician data. After reviewing the method Mr. Lensmeyer determined that there were wash cycles
that needed to be added to the method and those were added while we were there. The installation took a
day and a half to satisfy Mr. Lensmeyer’s approval of the install with a half day required only to replace
a Fast Wash Station that was thought to be defective.
2
During the first month Mr. Lensmeyer ran samples using a combination of control and patient samples
to optimize the method. During this time we provided per his request 10mg, then 3, 5, and 15mg
bedmass weights to see which weight actually performed best. 15mg bedmass was determined to be his
optimum volume.
Phase II was to actually test the device in a real time situation. Because of the positive results of the
initial data, Mr. Lensmeyer asked for and received permission to run samples in-line during limited
periods of phase II. In preparation for the event the PAL was installed on a metal lab cart so that the
system could be plumbed during the research period and removed on a daily basis to perform real-time
samples. Unfortunately, Mr. Lensmeyer’s LC/MS system had an antiquated version of ChemStation and
the installation was not a success.
In the report Mr. Lensmeyer reports that there was a hardware issue during the end of the research. It
appeared that the Fast Wash Station had been bumped and the X, Y axis disturbed to the extent that Z
axis error messages occurred. While we worked out one location over the phone with him there were
actually three locations that were affected. Mr. Lensmeyer believed that had enough data to write his
report and discontinued use of the PAL. It is our belief that had the PAL been installed in a proper
setting the automation would have been robust.
From here the report speaks for itself regarding the data however there are some hardware conclusions
that we believed needed additional investigation.
OpAns, LLC Continuation of the Research: Mr. Lensmeyer’s method was performed on an Agilent
1100 LC/MS system. Because of the sensitivity, Mr. Lensmeyer’s method requires a 1mL syringe to
prep samples for his lab while his injection volume was only 5uL’s, out of the range of the 1mL
syringe’s 100uL minimum injection. In a real-world application this could be overcome by eluting into
an open well and mixing steps could be utilized to make sure the 5uL injection had adequate amounts of
the desired compounds. While you can overfill the sample loop to use this syringe, the time that was
required was not as desirable as offering an injection that was closer to the injection volume of the
method.
A more expensive solution to this is one that occurs in GC applications. In this instance the requirement
of a second rail on the CTC PAL to offer a syringe in a sufficient volume to mange the small injection
could be installed on the existing system.
OpAns, LLC has been asked to review the data and concluded that LC/MS/MS could be utilized to scale
the application down to at least the range of a 100uL syringe. This work is in process at the moment. Dr.
Eaton, coming from an equivalent lab in Texas believes (as do we) that any new instrument being
purchased by a lab would set a minimum standard of LC/MS/MS. While OpAns was provided the Strata
SDB-L 15mg devices to produce new baseline data, they also believe that LC/MS/MS applications
could use more mainstream sorbents such as C8 or CN to scale the application down to the 100uL target.
Once this data is complete we will combine the data into one report that we believe will give a wide
range of options to choose from when converting to ITSP from existing applications.
Abstracts have been submitted to ASMS for possible publication at the Annual Meeting.
3
Development and Validation of a Method for Extraction of Immunosuppressive drugs from Whole
Blood using the MicroLiter ITSP extraction device on a CTC Analytics PAL
Introduction
Cyclosporine, sirolimus and tacrolimus are fungal-based drugs that are used extensively to suppress
rejection of transplanted organs in patients. These drugs have a narrow therapeutic window and thereby
require close monitoring of blood concentrations to achieve effective levels and to prevent toxicity. In
2004 we developed and published (Ther Drug Monit 2004; vol. 27(1): pp 1-7) a liquid
chromatography/mass spectrometry (LC/MS) method (See Appendix A) for the three drugs. A solid-
phase, semi-automated extraction process on the Gilson XL4 was incorporated prior to LC/MS analysis.
In this report, we describe an alternative way of extracting the three drugs from blood. Here,
MicroLiter’s ITSP with CTC Analytics PAL handler was used in place of the Gilson XL4. The
extraction protocol was essentially the same for both extraction units except, the ITSP used smaller
reagent volumes and a different sorbent weight of SDBL (Phenomenex).
Procedure
Following the protocol described in the published method, we precipitated whole blood proteins with
750 uL of precipitating reagent and 250 uL whole blood (EDTA). After a 10 minute incubation at room
temperature, the sample was centrifuged (10 min). The clear supernatant was decanted into a test tube
containing 500 uL water. This mixture was placed on the ITSP/PAL unit. Briefly, the ITSP probe picks
up a sorbent cartridge (SDBL, 15 mg), places it on the dock and primes the sorbent with 0.50 mL
acetonitrile followed by 0.50 mL 10% acetonitrile in water. A 1.0 mL portion of the diluted sample
supernatant is picked up by the probe/syringe and applied to the sorbent cartridge. The sorbent cartridge
is washed with 0.50 mL of 30% acetonitrile in water. The probe/syringe is rinsed with 0.50 mL of 10%
acetonitrile in water. Next, the probe/syringe loads 0.6 mL of eluting solution (acetonitrile) and then
picks up the sorbent cartridge containing the retained drugs. The probe/syringe with attached sorbent
cartridge moves to the elution area and forces the liquid through the cartridge into an empty vial. An
aliquot of the eluate is injected into the LC/MS for analysis. In this study, we placed the vial with eluate
directly onto the sample tray of the LC/MS. The program details of the ITSP/PAL along with typical
LC/MS chromatograms are located in Appendix B.
4
Studies and Results
With-run Precision
Three whole-blood control products (Utak 2, Utak 3, BioRad Lyphochek 1) that contain cyclosporine,
tacrolimus and sirolimus in a range of concentrations were assayed in single run in replicate (n=20).
Results are listed in Appendix C. In summary the final results were:
Lyphochek 1 Tacrolimus Sirolimus Cyclosporine
Mean ng/mL 3.49 4.03 73.69
Std. Dev. ng/ml 0.13 0.23 1.39
CV (%) 3.6 % 5.7 % 1.9 %
Utak 2
Mean ng/mL 15.07 18.09 515.5
Std. Dev. ng/ml 0.38 0.49 5.19
CV (%) 2.5 % 2.7 % 1.0 %
Utak 3
Mean ng/mL 22.52 28.19 1196.3
Std. Dev. ng/ml 0.46 1.01 13.29
CV (%) 2.1 % 3.6% 1.1 %
Conclusion. Results are excellent. Generally CV’s < 10% are acceptable.
Between-Run Precision
Three whole-blood control products (Utak 2, Utak 3, BioRad Lyphochek 1) that contain cyclosporine,
tacrolimus and sirolimus in a range of concentrations were assayed in 20 different analytical runs for
each control product. Results are listed in Appendix D. In summary the final results were:
ITSP:
Lyphochek1 Tacrolimus Sirolimus Cyclosporine
Mean ng/mL 3.63 3.86 78.22
Std. Dev. ng/ml 0.14 0.30 3.34
CV (%) 3.8 % 7.8 % 4.3 %
Utak 2
Mean ng/mL 15.45 18.47 526.8
Std. Dev. ng/ml 0.29 0.63 6.82
CV (%) 1.9 % 3.4 % 1.3 %
Utak 3
Mean ng/mL 23.56 29.85 1228.63
Std. Dev. ng/ml 0.58 1.15 20.02
CV (%) 2.5 % 3.8 % 1.6 %
5
Here are between-run precision for the Gilson XL4 extraction
Lyphochek 1 Tacrolimus Sirolimus Cyclosporine
Mean ng/mL 3.54 3.70 72.13
Std. Dev. ng/ml 0.153 0.263 1.94
CV (%) 4.3 % 7.1% 2.7 %
Utak 2
Mean ng/mL 15.11 18.59 526.8
Std. Dev. ng/ml 0.463 0.895 10.47
CV (%) 3.1 % 4.8 % 1.9 %
Utak 3
Mean ng/mL 23.32 30.01 1231.1
Std. Dev. ng/ml 0.839 1.909 28.76
CV (%) 3.6 % 6.3 % 2.33 %
Conclusion. Results are excellent. Generally CV’s < 10% are acceptable.
Linearity
Drug-free whole blood was supplemented with a range of concentrations of the three drugs. The samples
were assayed and graphs of the analytical result versus the supplemented concentration were created.
This study was performed in two separate runs on different days with a different set of concentrations
each day. Data is listed and results are listed in Appendix E.
Composite graphs of the data are presented here:
6
Sirolimus Linearity
0
20
40
60
80
100
120
140
0.0 20.0 40.0 60.0 80.0 100.
0
120.
0
140.
0
Conc Sirolimus Added to Whole Blood ng/mL
AnalyticalResultng/mL
Tacrolimus Linearity
0
10
20
30
40
50
60
70
80
0.0 20.0 40.0 60.0 80.0
Conc Tacrolimus Added to Whole Blood
ng/mL
AnalyticalResultng/mL
Cyclosporine Linearity
0
500
1000
1500
2000
2500
3000
3500
0 500 1000 1500 2000 2500 3000 3500
Conc CsA Added to Whole Blood ng/mL
AnalyticalResultng/mL
Sirolimus
n = 10
r = 0.9997
y = 1.00x + .005 ng/mL
Sy/x = 1.00 ng/mL
Tacrolimus
n = 10
r = 0.9996
y = 1.00x + .014 ng/mL
Sy/x = 0.611 ng/mL
Cyclosporine
n = 10
r = 1.0000
y = 0.999x + 0.245 ng/mL
Sy/x = 0.512 ng/mL
7
Gilson XL4 Linearity
Sirolimus Tacrolimus Cyclosporine
n = 12 n = 12 n =12
r = 0.9999 r = 0.9998 r = 0.9998
y = 1.01x + .04 ng/mL y = 1.007x + 0.26 ng/ml y = 0.993x + 4.02
Linear from 1 to Linear from 1 to Linear from 25
at least 80 ng/mL at least 80 ng/mL ng/mL to at least
2000 ng/mL
Conclusion: Linearity is acceptable when graphs are evaluated visually and regression data are
examined.
Patient Sample Comparisons
Samples from patients receiving the immunosuppressive drugs and drug-supplemented samples were
extracted by both the ITSP/PAL and the Gilson XL4. Results were compared and regression graphs
were created. Analytical results can be found in Appendix F.
Composite graphs of the data are presented here:
8
Sirolimus Comparisons
0
10
20
30
40
50
60
70
80
90
100
0 20 40 60 80 100
Gilson Extraction Conc ng/mL
ITSPExtractionConcng/mL
Tacrolimus Comparisions
0
10
20
30
40
50
60
70
80
90
0 10 20 30 40 50 60 70 80 90
Gilson Extraction Conc ng/mL
ITSPExtractionConcng/mL
Cyclosporine Comparisons
0
500
1000
1500
2000
2500
0 500 1000 1500 2000 2500
Gilson Extraction Conc ng/mL
ITSPExtractionConcng/mL
Sirolimus
n = 23
r = 0.9982
y = 1.08x – 1.38 ng/mL
Sy/x = 1.51 ng/mL
Tacrolimus
n = 25
r = 0.9983
y = 0.972x + 0.24 ng/mL
Sy/x = 1.01 ng/mL
Cyclosporine
n = 23
r = 0.9981
y = 1.01x + 1.71 ng/mL
Sy/x = 37.6 ng/mL
9
Conclusion:
Comparison results are excellent. Results for extractions with Gilson XL4 and the ITSP/PAL are
essentially identical.
Low Limit of Quantization (LLQ)
The lyphochek 1 control was diluted with drug-free whole blood to achieve concentrations of
approximately 0-1 ng/mL for Tacrolimus and Sirolimus. For cyclosporine concentrations were 10-20
ng/mL for cyclosporine. Each dilution was assayed in replicate (n = 9).
Results:
LLQ is for cyclosporine is 8 -10 ng/mL and for tacrolimus 0.8 – 1.0 ng/mL and sirolimus 1.5 – 1.8
ng/mL. Lower limits of LLQ can be achieved by injecting larger sample volumes.
Conclusion: LLQ for the ITSP extraction and Gilson XL4 extractions are essentially the same.
Comments
1. The operation of the ITSP/PAL instrument for this study was essentially trouble-free. On two
occasions, the Z -axis of the probe/syringe became misaligned when a liquid was about to be
applied to the sorbent cartridge setting in the dock and when the probe/syringe was attempting
dock on the port for the first priming solution. The problem was corrected by optimizing the
setting via the PAL software. Both issues occurred during the last week of the study. Also, at
times when it was necessary to abort a run, the computer program failed (lost address of
probe/syringe location) when a sequence was re-initiated. Rebooting the computer corrected the
problem.
2. The benefits of the ITSP/PAL over the Gilson XL4 for the extraction of the immunosuppressive
drugs:
> 60% decrease in solvent usage
decrease in amount of disposable glassware required
less amount of sorbent (SDBL) required (15 vs. 25 mg)
less manual intervention necessary/less labor
can directly inject into LC/MS (This study did not use this feature)
“just-in-time” extraction allows easy insertion of stat samples
potentially allows just one technologist to process all samples for the three drugs that come
into our laboratory during day shift. (Improvement)

More Related Content

PDF
Pain management-brochure itsp
Mark Hayward
 
PPTX
Presentation 6
ACCULAB USA
 
PPT
Automated analysis by yatin sankharva copy
Patel Parth
 
PPTX
Automation
RGCL
 
PPTX
Automation
Tapeshwar Yadav
 
PDF
Planning Considerations in Total Automation of Clinical Laboratory
Dr.Eng. Walid Tarawneh
 
PPTX
Automation in Biochemistry (Autoanalyzers)
Pradeep Singh Narwat
 
PPTX
Automation in the Clinical Lab
Ali Raza Ph.D
 
Pain management-brochure itsp
Mark Hayward
 
Presentation 6
ACCULAB USA
 
Automated analysis by yatin sankharva copy
Patel Parth
 
Automation
RGCL
 
Automation
Tapeshwar Yadav
 
Planning Considerations in Total Automation of Clinical Laboratory
Dr.Eng. Walid Tarawneh
 
Automation in Biochemistry (Autoanalyzers)
Pradeep Singh Narwat
 
Automation in the Clinical Lab
Ali Raza Ph.D
 

What's hot (11)

PPTX
Full automation
jdyjdo
 
PDF
ISES 2015 Max Mascelloni Cell exposure
Massimiliano Mascelloni
 
PDF
Auto analyzer
Tapeshwar Yadav
 
PPTX
Dual-Enzyme CatalyCEST
Gary Tyree
 
PDF
A novel approach in identification of blood group using laser technology
eSAT Publishing House
 
PDF
Improvement of bacterial estimation in urinalysis by iQ®200 ELITE®: capture i...
kridsada31
 
PDF
dry blood spotting technique,DBS
Vishal N.
 
PPTX
poster v5 (1)
Wilfried Tadjou
 
PPTX
Application of Computational and High-Throughput in vitro Screening for Prior...
U.S. EPA Office of Research and Development
 
PDF
Abbott HCV-HemaSpot final 0424
Shelley Hossenlopp
 
Full automation
jdyjdo
 
ISES 2015 Max Mascelloni Cell exposure
Massimiliano Mascelloni
 
Auto analyzer
Tapeshwar Yadav
 
Dual-Enzyme CatalyCEST
Gary Tyree
 
A novel approach in identification of blood group using laser technology
eSAT Publishing House
 
Improvement of bacterial estimation in urinalysis by iQ®200 ELITE®: capture i...
kridsada31
 
dry blood spotting technique,DBS
Vishal N.
 
poster v5 (1)
Wilfried Tadjou
 
Application of Computational and High-Throughput in vitro Screening for Prior...
U.S. EPA Office of Research and Development
 
Abbott HCV-HemaSpot final 0424
Shelley Hossenlopp
 
Ad

Viewers also liked (9)

PPTX
Immunopharmacology
Manish Kumar
 
PPTX
immunosuppression drugs
Aniket Narkar
 
PPT
Immunopharmacology
alaa essa
 
PPTX
Class immunosuppressants 2
Raghu Prasada
 
PPTX
Immunosuppressants [autosaved]
B.Devadatha datha
 
PPT
Immunopharmacology
Jincy13
 
PPTX
Immunomodulators(VK)
Dr. Abhavathi Vijay Kumar
 
PPT
11.IMMUNOSUPPRESSANTS
Saminathan Kayarohanam
 
PPTX
Immunosuppressants
Zulcaif Ahmad
 
Immunopharmacology
Manish Kumar
 
immunosuppression drugs
Aniket Narkar
 
Immunopharmacology
alaa essa
 
Class immunosuppressants 2
Raghu Prasada
 
Immunosuppressants [autosaved]
B.Devadatha datha
 
Immunopharmacology
Jincy13
 
Immunomodulators(VK)
Dr. Abhavathi Vijay Kumar
 
11.IMMUNOSUPPRESSANTS
Saminathan Kayarohanam
 
Immunosuppressants
Zulcaif Ahmad
 
Ad

Similar to Immunosuppressants in Blood Article - MicroLiter (20)

PDF
DUI-D in Urine Poster
Rick Youngblood
 
PDF
Dui d in urine poster itsp
Mark Hayward
 
PDF
Drugs of Abuse in Post Mortem Blood Poster
Rick Youngblood
 
PDF
Journal Combinatorial Chemistry 2006 v8 p820
Peter Tidswell
 
PDF
EXAMINING THE EFFECT OF FEATURE SELECTION ON IMPROVING PATIENT DETERIORATION ...
IJDKP
 
PDF
Unified-Drug-Analysis_Hayward_HPLC-2016
Mark Hayward
 
PDF
Method Development Poster-MicroLiter
Rick Youngblood
 
PDF
Automated SPE-LC/MS/MS method development using ITSP
Mark Hayward
 
PDF
H-Validation2 PPT
Andrew Keens
 
PDF
DMLT (2nd Year) : Automation in Clinical Biochemistry and POCT - Some basic c...
PrabhatNigam6
 
PPT
Solaris bio cellular analysis
zmusial
 
PDF
Identifying and Evaluating Usability Heuristics Applicable to Clinical Labora...
Alexandre Savaris
 
PPTX
SURE poster
Dana Joaquin
 
PPTX
Bioanalytical validation personal perspective
E. Dennis Bashaw
 
PDF
Chromatographic Analysis of Pharmaceuticals Second Edition by John A. Adamovics
Rohit K.
 
PPTX
Automatic And Semi Automatic Analyser Biochemistry
MdShamsTabrez4
 
PPTX
CoMPARA: Collaborative Modeling Project for Androgen Receptor Activity
Kamel Mansouri
 
PDF
Analytical Considerations When Monitoring Pain Medications by LC-MS/MS
David Masters-Moore
 
PDF
5-MEO-DALT produces short-lived psychedelic effects
BRC SHOP
 
PDF
pnl_05_en_006_web
Dr. Christoph S. Eberle, MICR
 
DUI-D in Urine Poster
Rick Youngblood
 
Dui d in urine poster itsp
Mark Hayward
 
Drugs of Abuse in Post Mortem Blood Poster
Rick Youngblood
 
Journal Combinatorial Chemistry 2006 v8 p820
Peter Tidswell
 
EXAMINING THE EFFECT OF FEATURE SELECTION ON IMPROVING PATIENT DETERIORATION ...
IJDKP
 
Unified-Drug-Analysis_Hayward_HPLC-2016
Mark Hayward
 
Method Development Poster-MicroLiter
Rick Youngblood
 
Automated SPE-LC/MS/MS method development using ITSP
Mark Hayward
 
H-Validation2 PPT
Andrew Keens
 
DMLT (2nd Year) : Automation in Clinical Biochemistry and POCT - Some basic c...
PrabhatNigam6
 
Solaris bio cellular analysis
zmusial
 
Identifying and Evaluating Usability Heuristics Applicable to Clinical Labora...
Alexandre Savaris
 
SURE poster
Dana Joaquin
 
Bioanalytical validation personal perspective
E. Dennis Bashaw
 
Chromatographic Analysis of Pharmaceuticals Second Edition by John A. Adamovics
Rohit K.
 
Automatic And Semi Automatic Analyser Biochemistry
MdShamsTabrez4
 
CoMPARA: Collaborative Modeling Project for Androgen Receptor Activity
Kamel Mansouri
 
Analytical Considerations When Monitoring Pain Medications by LC-MS/MS
David Masters-Moore
 
5-MEO-DALT produces short-lived psychedelic effects
BRC SHOP
 

More from Rick Youngblood (20)

PDF
Automated Chromatographic Solid-Phase Extraction Using an Autosampler _ Ameri...
Rick Youngblood
 
PPSX
ITSP PAL Training - Intermediate
Rick Youngblood
 
PPSX
ITSP PAL Training - Advanced
Rick Youngblood
 
PPSX
ITSP PAL Training - Basic
Rick Youngblood
 
PPSX
ITSP Dual Rail Project
Rick Youngblood
 
PPSX
ITSP New Stuff
Rick Youngblood
 
PPSX
Nova Analytica
Rick Youngblood
 
PPSX
Optimization of Automated Online SPE-LC-MS
Rick Youngblood
 
PPSX
Unified-Drug-Analysis-PittCon-2016
Rick Youngblood
 
PPSX
ITSP Setup on PAL-xt
Rick Youngblood
 
PPSX
ITSP Introduction
Rick Youngblood
 
PDF
Barbiturates and THCA in Urine at OpAns
Rick Youngblood
 
PDF
Androstenedione and Testosterone in Serum
Rick Youngblood
 
PDF
Manual Single Cartridge Holder Flyer
Rick Youngblood
 
PDF
Streptomycin in Honey Poster
Rick Youngblood
 
PDF
Automated SPE for Capillary Microsampling Poster
Rick Youngblood
 
PDF
as119
Rick Youngblood
 
PDF
QuEChERS Pittcon 2014 Poster
Rick Youngblood
 
PDF
Drugs_of_Abuse_Application_Note
Rick Youngblood
 
PDF
Go Green Brochure-MicroLiter
Rick Youngblood
 
Automated Chromatographic Solid-Phase Extraction Using an Autosampler _ Ameri...
Rick Youngblood
 
ITSP PAL Training - Intermediate
Rick Youngblood
 
ITSP PAL Training - Advanced
Rick Youngblood
 
ITSP PAL Training - Basic
Rick Youngblood
 
ITSP Dual Rail Project
Rick Youngblood
 
ITSP New Stuff
Rick Youngblood
 
Nova Analytica
Rick Youngblood
 
Optimization of Automated Online SPE-LC-MS
Rick Youngblood
 
Unified-Drug-Analysis-PittCon-2016
Rick Youngblood
 
ITSP Setup on PAL-xt
Rick Youngblood
 
ITSP Introduction
Rick Youngblood
 
Barbiturates and THCA in Urine at OpAns
Rick Youngblood
 
Androstenedione and Testosterone in Serum
Rick Youngblood
 
Manual Single Cartridge Holder Flyer
Rick Youngblood
 
Streptomycin in Honey Poster
Rick Youngblood
 
Automated SPE for Capillary Microsampling Poster
Rick Youngblood
 
QuEChERS Pittcon 2014 Poster
Rick Youngblood
 
Drugs_of_Abuse_Application_Note
Rick Youngblood
 
Go Green Brochure-MicroLiter
Rick Youngblood
 

Immunosuppressants in Blood Article - MicroLiter

  • 1. 1 University of Wisconsin Medical Center Research for Determination of Immunosuppressants in Whole Blood Background: Gary Lensmeyer of the University of Wisconsin Medical Center (UWMC) is responsible for exploring new technologies for his laboratory. During Monica Howland’s (MicroLiter) employment with 3M she worked with Mr. Lensmeyer to evaluate and assist in development of the Empore line of SPE membranes. MicroLiter contracted with UMWC to evaluate Instrument Top Sample Prep (ITSP) by developing a method that would typically be used in his lab. UWMC specializes in transplantation of organs. Immunosuppressants, prescription pain killers, and Vitamin D were the principal high volume assays that were run daily in his lab. Mr. Lensmeyer chose Immunosuppressants as the method he would transfer to our ITSP. Terms of the UWMC Contract: Mr. Lensmeyer was asked to independently evaluate ITSP to give us good, bad and indifferent opinions relating to the device. While we thought he would have a positive experience we did not ask him to produce data that offered only the benefit of the device. Because of conflicts in scheduling MicroLiter’s HTC PAL, the research was conducted in two phases. The first phase would be development of the method on the HTC PAL. The second phase would be to run the device against current laboratory operations. Due to the critical nature of the work being performed in his lab, the Lab Manager, Anita Iwanski, would not allow the instrument to be attached on-line with the Agilent LC/MS that was currently running the assay. Mrs. Anita Iwanski, Lab Manager made it clear that there was no money available for purchase of the system until July 2010. The Work Product: Mr. Lensmeyer’s current method utilized Phenomenex Strata SDB-L sorbents and we were asked to use this material for a true apples-to-apples comparison. We asked for and obtained a sufficient quantity to begin our study. Initially 10mg bedmass weights were prepared. Mr. Lensmeyer was given 100 devices in advance of the delivery of the instrument to try and produce a manual method that we could program into the PAL. He was able to recreate the method, supplied us the conditions and we wrote the automation in advance of the delivery of the instrument. On the day we installed the instrument, Mr. Lensmeyer immediately prepared solvents and supernatants and our first test run of the instrument prepped samples that gave Mr. Lensmeyer baseline data with which to start. The very first sample produced data that would have been acceptable to the lab to provide the physician data. After reviewing the method Mr. Lensmeyer determined that there were wash cycles that needed to be added to the method and those were added while we were there. The installation took a day and a half to satisfy Mr. Lensmeyer’s approval of the install with a half day required only to replace a Fast Wash Station that was thought to be defective.
  • 2. 2 During the first month Mr. Lensmeyer ran samples using a combination of control and patient samples to optimize the method. During this time we provided per his request 10mg, then 3, 5, and 15mg bedmass weights to see which weight actually performed best. 15mg bedmass was determined to be his optimum volume. Phase II was to actually test the device in a real time situation. Because of the positive results of the initial data, Mr. Lensmeyer asked for and received permission to run samples in-line during limited periods of phase II. In preparation for the event the PAL was installed on a metal lab cart so that the system could be plumbed during the research period and removed on a daily basis to perform real-time samples. Unfortunately, Mr. Lensmeyer’s LC/MS system had an antiquated version of ChemStation and the installation was not a success. In the report Mr. Lensmeyer reports that there was a hardware issue during the end of the research. It appeared that the Fast Wash Station had been bumped and the X, Y axis disturbed to the extent that Z axis error messages occurred. While we worked out one location over the phone with him there were actually three locations that were affected. Mr. Lensmeyer believed that had enough data to write his report and discontinued use of the PAL. It is our belief that had the PAL been installed in a proper setting the automation would have been robust. From here the report speaks for itself regarding the data however there are some hardware conclusions that we believed needed additional investigation. OpAns, LLC Continuation of the Research: Mr. Lensmeyer’s method was performed on an Agilent 1100 LC/MS system. Because of the sensitivity, Mr. Lensmeyer’s method requires a 1mL syringe to prep samples for his lab while his injection volume was only 5uL’s, out of the range of the 1mL syringe’s 100uL minimum injection. In a real-world application this could be overcome by eluting into an open well and mixing steps could be utilized to make sure the 5uL injection had adequate amounts of the desired compounds. While you can overfill the sample loop to use this syringe, the time that was required was not as desirable as offering an injection that was closer to the injection volume of the method. A more expensive solution to this is one that occurs in GC applications. In this instance the requirement of a second rail on the CTC PAL to offer a syringe in a sufficient volume to mange the small injection could be installed on the existing system. OpAns, LLC has been asked to review the data and concluded that LC/MS/MS could be utilized to scale the application down to at least the range of a 100uL syringe. This work is in process at the moment. Dr. Eaton, coming from an equivalent lab in Texas believes (as do we) that any new instrument being purchased by a lab would set a minimum standard of LC/MS/MS. While OpAns was provided the Strata SDB-L 15mg devices to produce new baseline data, they also believe that LC/MS/MS applications could use more mainstream sorbents such as C8 or CN to scale the application down to the 100uL target. Once this data is complete we will combine the data into one report that we believe will give a wide range of options to choose from when converting to ITSP from existing applications. Abstracts have been submitted to ASMS for possible publication at the Annual Meeting.
  • 3. 3 Development and Validation of a Method for Extraction of Immunosuppressive drugs from Whole Blood using the MicroLiter ITSP extraction device on a CTC Analytics PAL Introduction Cyclosporine, sirolimus and tacrolimus are fungal-based drugs that are used extensively to suppress rejection of transplanted organs in patients. These drugs have a narrow therapeutic window and thereby require close monitoring of blood concentrations to achieve effective levels and to prevent toxicity. In 2004 we developed and published (Ther Drug Monit 2004; vol. 27(1): pp 1-7) a liquid chromatography/mass spectrometry (LC/MS) method (See Appendix A) for the three drugs. A solid- phase, semi-automated extraction process on the Gilson XL4 was incorporated prior to LC/MS analysis. In this report, we describe an alternative way of extracting the three drugs from blood. Here, MicroLiter’s ITSP with CTC Analytics PAL handler was used in place of the Gilson XL4. The extraction protocol was essentially the same for both extraction units except, the ITSP used smaller reagent volumes and a different sorbent weight of SDBL (Phenomenex). Procedure Following the protocol described in the published method, we precipitated whole blood proteins with 750 uL of precipitating reagent and 250 uL whole blood (EDTA). After a 10 minute incubation at room temperature, the sample was centrifuged (10 min). The clear supernatant was decanted into a test tube containing 500 uL water. This mixture was placed on the ITSP/PAL unit. Briefly, the ITSP probe picks up a sorbent cartridge (SDBL, 15 mg), places it on the dock and primes the sorbent with 0.50 mL acetonitrile followed by 0.50 mL 10% acetonitrile in water. A 1.0 mL portion of the diluted sample supernatant is picked up by the probe/syringe and applied to the sorbent cartridge. The sorbent cartridge is washed with 0.50 mL of 30% acetonitrile in water. The probe/syringe is rinsed with 0.50 mL of 10% acetonitrile in water. Next, the probe/syringe loads 0.6 mL of eluting solution (acetonitrile) and then picks up the sorbent cartridge containing the retained drugs. The probe/syringe with attached sorbent cartridge moves to the elution area and forces the liquid through the cartridge into an empty vial. An aliquot of the eluate is injected into the LC/MS for analysis. In this study, we placed the vial with eluate directly onto the sample tray of the LC/MS. The program details of the ITSP/PAL along with typical LC/MS chromatograms are located in Appendix B.
  • 4. 4 Studies and Results With-run Precision Three whole-blood control products (Utak 2, Utak 3, BioRad Lyphochek 1) that contain cyclosporine, tacrolimus and sirolimus in a range of concentrations were assayed in single run in replicate (n=20). Results are listed in Appendix C. In summary the final results were: Lyphochek 1 Tacrolimus Sirolimus Cyclosporine Mean ng/mL 3.49 4.03 73.69 Std. Dev. ng/ml 0.13 0.23 1.39 CV (%) 3.6 % 5.7 % 1.9 % Utak 2 Mean ng/mL 15.07 18.09 515.5 Std. Dev. ng/ml 0.38 0.49 5.19 CV (%) 2.5 % 2.7 % 1.0 % Utak 3 Mean ng/mL 22.52 28.19 1196.3 Std. Dev. ng/ml 0.46 1.01 13.29 CV (%) 2.1 % 3.6% 1.1 % Conclusion. Results are excellent. Generally CV’s < 10% are acceptable. Between-Run Precision Three whole-blood control products (Utak 2, Utak 3, BioRad Lyphochek 1) that contain cyclosporine, tacrolimus and sirolimus in a range of concentrations were assayed in 20 different analytical runs for each control product. Results are listed in Appendix D. In summary the final results were: ITSP: Lyphochek1 Tacrolimus Sirolimus Cyclosporine Mean ng/mL 3.63 3.86 78.22 Std. Dev. ng/ml 0.14 0.30 3.34 CV (%) 3.8 % 7.8 % 4.3 % Utak 2 Mean ng/mL 15.45 18.47 526.8 Std. Dev. ng/ml 0.29 0.63 6.82 CV (%) 1.9 % 3.4 % 1.3 % Utak 3 Mean ng/mL 23.56 29.85 1228.63 Std. Dev. ng/ml 0.58 1.15 20.02 CV (%) 2.5 % 3.8 % 1.6 %
  • 5. 5 Here are between-run precision for the Gilson XL4 extraction Lyphochek 1 Tacrolimus Sirolimus Cyclosporine Mean ng/mL 3.54 3.70 72.13 Std. Dev. ng/ml 0.153 0.263 1.94 CV (%) 4.3 % 7.1% 2.7 % Utak 2 Mean ng/mL 15.11 18.59 526.8 Std. Dev. ng/ml 0.463 0.895 10.47 CV (%) 3.1 % 4.8 % 1.9 % Utak 3 Mean ng/mL 23.32 30.01 1231.1 Std. Dev. ng/ml 0.839 1.909 28.76 CV (%) 3.6 % 6.3 % 2.33 % Conclusion. Results are excellent. Generally CV’s < 10% are acceptable. Linearity Drug-free whole blood was supplemented with a range of concentrations of the three drugs. The samples were assayed and graphs of the analytical result versus the supplemented concentration were created. This study was performed in two separate runs on different days with a different set of concentrations each day. Data is listed and results are listed in Appendix E. Composite graphs of the data are presented here:
  • 6. 6 Sirolimus Linearity 0 20 40 60 80 100 120 140 0.0 20.0 40.0 60.0 80.0 100. 0 120. 0 140. 0 Conc Sirolimus Added to Whole Blood ng/mL AnalyticalResultng/mL Tacrolimus Linearity 0 10 20 30 40 50 60 70 80 0.0 20.0 40.0 60.0 80.0 Conc Tacrolimus Added to Whole Blood ng/mL AnalyticalResultng/mL Cyclosporine Linearity 0 500 1000 1500 2000 2500 3000 3500 0 500 1000 1500 2000 2500 3000 3500 Conc CsA Added to Whole Blood ng/mL AnalyticalResultng/mL Sirolimus n = 10 r = 0.9997 y = 1.00x + .005 ng/mL Sy/x = 1.00 ng/mL Tacrolimus n = 10 r = 0.9996 y = 1.00x + .014 ng/mL Sy/x = 0.611 ng/mL Cyclosporine n = 10 r = 1.0000 y = 0.999x + 0.245 ng/mL Sy/x = 0.512 ng/mL
  • 7. 7 Gilson XL4 Linearity Sirolimus Tacrolimus Cyclosporine n = 12 n = 12 n =12 r = 0.9999 r = 0.9998 r = 0.9998 y = 1.01x + .04 ng/mL y = 1.007x + 0.26 ng/ml y = 0.993x + 4.02 Linear from 1 to Linear from 1 to Linear from 25 at least 80 ng/mL at least 80 ng/mL ng/mL to at least 2000 ng/mL Conclusion: Linearity is acceptable when graphs are evaluated visually and regression data are examined. Patient Sample Comparisons Samples from patients receiving the immunosuppressive drugs and drug-supplemented samples were extracted by both the ITSP/PAL and the Gilson XL4. Results were compared and regression graphs were created. Analytical results can be found in Appendix F. Composite graphs of the data are presented here:
  • 8. 8 Sirolimus Comparisons 0 10 20 30 40 50 60 70 80 90 100 0 20 40 60 80 100 Gilson Extraction Conc ng/mL ITSPExtractionConcng/mL Tacrolimus Comparisions 0 10 20 30 40 50 60 70 80 90 0 10 20 30 40 50 60 70 80 90 Gilson Extraction Conc ng/mL ITSPExtractionConcng/mL Cyclosporine Comparisons 0 500 1000 1500 2000 2500 0 500 1000 1500 2000 2500 Gilson Extraction Conc ng/mL ITSPExtractionConcng/mL Sirolimus n = 23 r = 0.9982 y = 1.08x – 1.38 ng/mL Sy/x = 1.51 ng/mL Tacrolimus n = 25 r = 0.9983 y = 0.972x + 0.24 ng/mL Sy/x = 1.01 ng/mL Cyclosporine n = 23 r = 0.9981 y = 1.01x + 1.71 ng/mL Sy/x = 37.6 ng/mL
  • 9. 9 Conclusion: Comparison results are excellent. Results for extractions with Gilson XL4 and the ITSP/PAL are essentially identical. Low Limit of Quantization (LLQ) The lyphochek 1 control was diluted with drug-free whole blood to achieve concentrations of approximately 0-1 ng/mL for Tacrolimus and Sirolimus. For cyclosporine concentrations were 10-20 ng/mL for cyclosporine. Each dilution was assayed in replicate (n = 9). Results: LLQ is for cyclosporine is 8 -10 ng/mL and for tacrolimus 0.8 – 1.0 ng/mL and sirolimus 1.5 – 1.8 ng/mL. Lower limits of LLQ can be achieved by injecting larger sample volumes. Conclusion: LLQ for the ITSP extraction and Gilson XL4 extractions are essentially the same. Comments 1. The operation of the ITSP/PAL instrument for this study was essentially trouble-free. On two occasions, the Z -axis of the probe/syringe became misaligned when a liquid was about to be applied to the sorbent cartridge setting in the dock and when the probe/syringe was attempting dock on the port for the first priming solution. The problem was corrected by optimizing the setting via the PAL software. Both issues occurred during the last week of the study. Also, at times when it was necessary to abort a run, the computer program failed (lost address of probe/syringe location) when a sequence was re-initiated. Rebooting the computer corrected the problem. 2. The benefits of the ITSP/PAL over the Gilson XL4 for the extraction of the immunosuppressive drugs: > 60% decrease in solvent usage decrease in amount of disposable glassware required less amount of sorbent (SDBL) required (15 vs. 25 mg) less manual intervention necessary/less labor can directly inject into LC/MS (This study did not use this feature) “just-in-time” extraction allows easy insertion of stat samples potentially allows just one technologist to process all samples for the three drugs that come into our laboratory during day shift. (Improvement)